GE
LOGIQ V5/LOGIQ V3
D
IRECTION
5496012-100, R
EVISION
3
B
ASIC
S
ERVICE
M
ANUAL
2 - 8
Section 2-2 - General Console Requirements
2-3-3
Minimal floor plan suggestion
CSI 8x10
Figure 2-1.
Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
Scale
:
Each square equals one square foot (app. 31 x 31 cm)
1. Door – at least 762 mm (30 inches)
2. Film
Viewer
3. Counter Top, Sink with hot and cold
water and Supplies Storage
4. Linen
Supply
5. Probes/Supplies
6. Examination Table – 1930 x 610 mm
(76 x 24 inches)
7. Footswitch
8. Stool
9. Ultrasound
system
10. External Peripherals
11. Dedicated Power Outlet - Circuit
Breaker protected and easily
accessible
12. Network Interface
13. 457 mm (18 inches) distance of Ultrasound system
from wall or objects
14. GE Cabinet for Software and Manuals
4
1
3
11
6
13
8
10
5
7
9
12
14
2
Содержание LOGIQ V3
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3 ...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO ...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks ...
Страница 145: ......
Страница 169: ......
Страница 170: ... 2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA ...