Chapter 4 - Functional Checks
4 - 21
CF/PDI Auto
Sample Volume
Yes
Set the default value at Utility -> Imaging -> CF Mode.
CF/PDI Focus
Depth
Yes
CF/PDI Frequency
Yes
CF/PDI Auto
Frequency
Yes
CF/PDI Center
Depth
Yes
PDI
Yes
Power Doppler Imaging (PDI) is a color flow mapping
technique used to map the strength of the Doppler signal
coming from the tissue rather than the frequency shift of
the signal. Using this technique, the ultrasound system
plots color flow based on the number of reflectors that are
moving, regardless of their velocity. PDI does not map
velocity, therefore it is not subject to aliasing.
TVI (Option)
Yes
Tissue Velocity Imaging (TVI) calculates and color-codes
the velocities in tissue. The tissue velocity information is
acquired by sampling of tissue Doppler velocity values at
discrete points.The information is stored in a combined
format with gray scale imaging during one or several
cardiac cycles with high temporal resolution.
TVD
Yes
TVD: Tissue Velocity Doppler: basing on TVI mode,
activate a sample volume of PW ventricular wall to get the
spectral information of the sample section.
Table 4-5
Color Flow Mode Controls
Control
Possible Bioeffect
Description/Benefit
Содержание LOGIQ V3
Страница 1: ...GE LOGIQ V5 LOGIQ V3 Basic Service Manual Part Number 5496012 100 Revision 3 ...
Страница 2: ......
Страница 11: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL ix ZH CN KO ...
Страница 27: ......
Страница 116: ...GE LOGIQ V5 LOGIQ V3 DIRECTION 5496012 100 REVISION 3 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks ...
Страница 145: ......
Страница 169: ......
Страница 170: ... 2014 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA ...