Overview
LOGIQ V2/LOGIQ V1
–
User Guide
1-13
Direction 5610736-100
Rev. 9
Battery
The lithium ion battery provides power when an AC power
source is not available. A battery in the battery bay is standard
with the LOGIQ V2/LOGIQ V1. Lithium ion batteries last longer
than conventional batteries and do not require replacement as
often. You can expect about 30 minutes of battery life with a
single fully charged battery in use to supply power to the
system.
NOTE:
While scanning with the battery supplying power only, the
battery life may be shorter. Always archive the data and keep
your attention on the battery status. When the battery power is
low, charge the battery immediately in case that scanning will be
interrupted and the data will be lost due to the automatic
shutdown of the system.
The lithium ion technology used in your system’s battery is
significantly less hazardous to the environment than the lithium
metal technology used in some other batteries (such as watch
batteries). Used batteries should not be placed with common
household waste products. Contact local authorities for the
location of a chemical waste collection program nearest you.
NOTE:
The battery is designed to work with LOGIQ V2/LOGIQ V1
systems only. Only use the batteries authorized by GE.
Temperature Requirements
The battery should be charged, discharged and stored within the
parameters outlined below:
• Operating
temperature:
•
Charge: 10 - 30°C (50 - 86°F).
•
Discharge: 10 - 40°C (50 - 104°F)
• Storage
temperature:
NOTE:
It is recommended that the battery remaining capacity
should be 40% ~ 60% when the battery storage begins.
•
Storage time < 3 months: -20 - 40°C (-4 - 104°F)
•
Storage time >= 3 months: -20 - 20°C (-4 - 68°F)
WARNING
Do not expose the battery to temperature over 60°C (140°F).
Keep it away from fire and other heat sources.
Содержание LOGIQ V2
Страница 4: ...i 2 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 This page intentionally left blank...
Страница 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 170: ...After the Exam is Over 3 8 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 Figure 3 10 Formats selection...
Страница 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 381: ......