Safety Precautions
LOGIQ V2/LOGIQ V1
–
User Guide
4-17
Direction 5610736-100
Rev. 9
EMC Performance (continued)
Portable and mobile radio communications equipment (e.g.
two-way radio, cellular/cordless telephones, wireless computer
networks) should be used no closer to any part of this system,
including cables, than determined according to the following
method:
Table 4-2: Portable and mobile radio communications equipment distance
requirements
Frequency Range:
150 kHz - 80 MHz
80 MHz - 800 MHz
800 MHz - 2.5 GHz
Calculation Method:
d=[3.5/V
1
] square root
of P
d = [3.5/E
1
] square root
of P
d = [7/E
1
] square root of
P
Where: d= separation distance in meters, P = rated power of the transmitter, V
1
=compliance value for
conducted RF, E
1
= compliance value for radiated RF
If the maximum
transmitter power in
watts is rated
The separation distance in meters should be
5
2.6
2.6
5.2
20
5.2
5.2
10.5
100
12.0
12.0
24.0
Содержание LOGIQ V2
Страница 4: ...i 2 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 This page intentionally left blank...
Страница 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 170: ...After the Exam is Over 3 8 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 Figure 3 10 Formats selection...
Страница 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 381: ......