Basic Measurements (continued)
Table 2-7: System Measurements and Accuracies
Measurement
Units
Useful Range
Accuracy
Limitations or Conditions
Depth
mm
Full Screen
<=10%
NA
Angle
degree
Full Screen
<=5%
NA
Distance:
Axial
mm
Full Screen
<5%
Linear Probes, Convex
Probes, Sector Probes
Lateral
mm
Full Screen
<5%
Linear Probes, Convex
Probes, Sector Probes
Circumference:
Trace
mm
Full Screen
<=10%
Linear Probes, Convex
Probes, Sector Probes
Ellipse
mm
Full Screen
<=5%
Linear Probes, Convex
Probes, Sector Probes
Area:
Trace
mm
2
Full Screen
<=5%
Linear Probes, Convex
Probes, Sector Probes
Ellipse
mm
2
Full Screen
<=5%
Linear Probes, Convex
Probes, Sector Probes
Time
s
Timeline Display
<5%
M mode, AM mode, CM mode,
PWD mode, CWD mode
Slope
mm/s
Timeline Display
<=10%
M mode, AM mode, CM mode,
Doppler SV
Position
mm
Full Screen
<=2 mm
PWD mode
Doppler Velocity
cm/s
Form 0 to 100 cm/s
From 100 to 130 cm/s
<15%
<10%
PWD mode, CWD mode
Doppler Angle
Correction
cm/s
From 0-80
°
<=5%
PWD mode
Содержание LOGIQ V2
Страница 4: ...i 2 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 This page intentionally left blank...
Страница 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 170: ...After the Exam is Over 3 8 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 Figure 3 10 Formats selection...
Страница 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 381: ......