Measurement and Analysis
LOGIQ V2/LOGIQ V1
–
User Guide
2-57
Direction 5610736-100
Rev. 9
TAMAX and TAMEAN
Manual Trace
The value measured depends upon the Vol Flow Method preset.
The two selections available are: Peak (TAMAX) and Mean
(TAMEAN).
To do a manual trace of TAMAX or TAMEAN:
1. Press
Measure
key to enter into measure mode, if
necessary.
Adjust
Caliper Change
in the Primary Menu to
S/D Trace
.
In the Utility -> Measure - Advanced preset menu set
Trace
preset to
Manual
.
2. To position the caliper at the trace start point, move the
Trackball
.
3. To fix the start point, press
Set
.
4. To trace the maximum values of the desired portion of the
spectrum, move the
Trackball
.
NOTE: To edit the trace line, move the
Trackball.
5. To complete the measurement, press
Set
. The system
displays the measurement values in the Results Window.