Quality Assurance
LOGIQ V2/LOGIQ V1
–
User Guide
3-65
Direction 5610736-100
Rev. 9
Functional resolution
Description
Functional resolution is an imaging system's ability to detect and
display the size, shape, and depth of an anechoic structure, as
opposed to a pin target. See Figure 3-35 for more information.
The very best possible image is somewhat less important than
reproducibility and stability over time. Routine tests at the same
settings should produce the same results.
Benefit
The data obtained will give a relative indication of the smallest
structure the system is capable of resolving at a given depth.
Procedure
To measure functional resolution:
1. Set the front panel TGC slide pots to their center (detent)
position.
2. Gain and acoustic output can be adjusted as necessary,
since these values are displayed on the monitor.
3. Scan a test phantom with a vertical row of anechoic cyst
targets to typical depths for the probe being used.
4. Evaluate the cysts at various depths for a good (round)
shape, well-defined borders and no fill in. Remember, TGC
slide pots are centered and should remain fixed. This may
NOT provide optimal cystic clearing.
5. Document all results for future reference and comparison.
Contact a Service Engineer if a greatly distorted image is
obtained.
Содержание LOGIQ V2
Страница 4: ...i 2 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 This page intentionally left blank...
Страница 8: ...i 6 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 92: ...Getting Started 1 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 170: ...After the Exam is Over 3 8 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9 Figure 3 10 Formats selection...
Страница 242: ...After the Exam is Over 3 80 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 288: ...Safety 4 46 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 380: ...Index 4 LOGIQ V2 LOGIQ V1 User Guide Direction 5610736 100 Rev 9...
Страница 381: ......