Probes and Biopsy
5-36
LOGIQ V2/LOGIQ V1
–
User Guide
Direction 5610736-100
Rev. 9
Multi Angle Biopsy Guide Assembly (continued)
3. Fit a convex piece of the biopsy bracket (a) into the concave
position of the probe (b).
Figure 5-11. Probe/Bracket Alignment
Hold the side (a) and tuck down the needle guide side (b)
until it clicks or locks in place.
Figure 5-12. Probe/Multi-angle Bracket Alignment 2
4. Place an adequate amount of coupling gel on the face of the
probe.