2277615-100
REV 0
g
GE Medical Systems
10/16
QA Reference Manual Section 4-2 Step 4.
from the skin line to the point furthermost in depth where echoes disappear and record the depth
in centimeters. The maximum depth through which the echographic pattern can be visualized
is: Less than 3.0
3.0
3.5
4.0
4.5
5.0
5.5
6.0
6.5
7.0
7.5
8.0
8.5
9.0
9.5
10.0
10.5
11.0
11.5
12.0
12.5
13.0 13.5 14.0 15.0
16.0 or higher
Did system penetration change since the previous test?
Change
No Change ( or First Test)
For Probe 2, adjust the system settings to produce the best possible overall image of the phantom.
This scan will be used to assess penetration capabilities, image uniformity, measurement
accuracy, axial resolution, lateral resolution and contrast detectability.
SECURE PAGE PRINT OF PROBE 2 SCAN
For scan repeatability during
the next QA check, place a
check mark in a box to
indicate the relative position of
each TCG control for the scan.
Gain: ________________________ Focus : ________________________
Dynamic Range: ____________________