GE MEDICAL SYSTEMS
LOGIQ
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS
2–42
REV 2
2–3–8 System Clock Adjust
Procedure
Follow these steps to adjust the system clock of the LOGIQ
a
200.
1.
Using the Arrow up/down keys, select the “System Clock Adjust” menu in the Top Menu. Refer to the
ILLUSTRATION 2–5.
2.
Press Set or Return key.
The System Clock Adjust Menu will appear on the monitor as shown in the ILLUSTRATION 2–31.
Service Utility MENU :
Top Menu
96/03/12
09:16:37
Current Software Version is 2.0
System Diagnosis
System Reset
Self Operation History
Option Set
Initialize :1 (I:America 2:Europe 3:Asia 4:Japan)
*System Clock Adjust: 98/3/27 15:10
Exit
DATABASE INITIALIZE MENU
ILLUSTRATION 2–31
3.
Input current date and time, and press Set or Return key.