GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--38
REV 0
2--3--2 System Diagnosis (Continued)
TEST NAME: MONITOR TEST
PURPOSE
To check monitor gradation ability, video circuit and video characteristics.
This test assists to check latitude (capability of density) of B/W printer.
CRITERIA
Refer to test pattern and gray scale shown on the monitor as in ILLUSTRATION 2--21. 16 gray scale
on the upper side of monitor. 16 gray scale on the lower side of monitor. The source of test pattern and
gray scale is graphic controller in MST and B_mode display memory in MVP, respectively
DISPLAY FORMAT OF RESULTS
145 mm
180 mm
MONITOR TEST IMAGE
ILLUSTRATION 2--21
Action if the test fails
Change the Monitor.
The following Assemblies are necessary for this test
RBF, TBF, SCAN, MVP, MST,Monitor