FUNCTIONAL CHECKS
4–3
REV 2
4–1
INTRODUCTION
This chapter provides procedures for quickly checking major functions of the LOGIQ
α
200 console, and SMPS
adjustments.
4–1–1
Required Equipment
To perform these tests, you’ll need a linear, or a convex transducer.
4–2
FUNCTIONAL CHECK PROCEDURES
4–2–1
Basic Controls
Step
Check
Expected Result
1
Connect the convex transducer
to ”Probe 1” connector.
2
Power On
After few seconds, the B mode screen should appears
as shown in ILLUSTRATION 4–1.
3
Rotate B/M Gain knob
Image gets lighter with CW rotation and darker with
CCW.
4
Press Map key to select another
gray scale Map.
The gray scale adjusts to each new Map selected.
5
Press Dyn Range Arrow up or
down key.
At lower Dynamic Range settings, image speckle
fades and prominent objects in the display are more
pronounced from the background image.
6
Rotate Depth knob.
The depth of image should be increased /decreased.
7
Slide TGC potentiometers (pots)
Image grows darker or brighter at depth equivalent to
pot’s location.
8
Press Zoom. key.
Press it again to exit.
The image should increase to X2 size.
9
Press Frame Avg key.
Image speckle fades and probe or wire movement is
smeared.
10
Press Edge Enhc key.
The edges inside the focal area(s) should become
lighter when you increase and darker as you decrease
its value.
11
Press Reverse key.
The image reverses the left/right orientation.
12
Press Reverse key again.
The image reverses again.
Содержание LOGIQ 200
Страница 4: ......
Страница 8: ......
Страница 10: ...05 23 00 MAC Page 2 of 2 ...
Страница 28: ...05 23 00 MAC Page 2 of 2 ...
Страница 87: ...LOGIQ α200 ...
Страница 88: ...LOGIQ 200 PRO ...
Страница 144: ......
Страница 168: ...GE MEDICAL SYSTEMS LOGIQ 200 PRO Series PROPRIETARY MANUAL 2242594 DIAGNOSTICS 2 31 B MODE CINE TEST ILLUSTRATION 2 15 ...
Страница 190: ......
Страница 196: ......