GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series SERVICE MANUAL
2235374
FUNCTIONAL CHECKS
4--10
REV 0
4--3--2
SMPS Adjustment Procedure
1.
Power LOGIQ 200 PRO Series ON. Wait for about 30 seconds to warm up the console.
2.
Connect a DVM to the appropriate place shown in Table 4--1.
3.
Verify that the voltages are as shown in Table 4--2.
TABLE 4-- 1
SMPS MEASUREMENT LOCATION
SMPS
MEASURE AT
RETURN AT
ADJUST AT
±
12V for ECG
ECG_12P, ECG_12M on
MOTHER3 BD ASSY
AGND3 (Ground) on
MOTHER3 BD ASSY
1 on SMPS ASSY
See ILLUSTRATION 4--7
+12 for Monitor
MON_12P on MOTHER3 BD
ASSY
AGND3 on MOTHER3 BD
ASSY
3 on SMPS ASSY
See ILLUSTRATION 4--7
+5V for Digital
VCC1 (+5V) on MVP BD ASSY
GND1 (GND) on MVP BD
ASSY
2 on SMPS ASSY
See ILLUSTRATION 4--7
TABLE 4--2
SMPS MEASUREMENT TOLERANCES
SMPS
MIN
MAX
--5V,
±
12V for Analog
--4.8 V
--12.6V
+12.6 V
--5.2 V
--11.4 V
+11.4 V
+12 for Monitor
+11.8 V
+12.2 V
+5V for Digital and Analog
+4.8 V
+5.2 V