GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--23
REV 0
2--3--2 System Diagnosis (Continued)
TEST NAME: Linear/Convex Probe Path Test
PURPOSE
To check a Convex/Linear Probe Path.
Overviews
--.Before starting this test, Test pattern for OQCARD should be loaded in the R data area of the test
probe.
Diagnosis
Condition
Probe Type
Linear Probe
Probe Name
Test Probe
Mode
Full B
Depth
4cm (TX scan ranges = 0 to 510)
Frame Rate
high
Freeze or Melt
Melt
Procedure
Diagnisis_LinearConvex-
PathTest
Save the current console parameters.
Change console parameters according to above.
outword (0x4c06, 0x02)
Wait for exit command (Ctrl+X).
Restore console parameters.
Result code
Checked
Decision
Check Image Patterns.
The following Assemblies are necessary for this test
BPHV, PROBE, RBF, TBF, SCAN, MVP, MSTE, Monitor