GE MEDICAL SYSTEMS
LOGIQ 200 PRO Series PROPRIETARY MANUAL
2242594
DIAGNOSTICS
2--34
REV 0
2--3--2 System Diagnosis (Continued)
TEST NAME: SYSTEM DRAM TEST
PURPOSE
To check DRAM in MSTE assy.
CRITERIA
Refer to result shown on the monitor.
Action if the test fails
Change MSTE board assy.
The following Assemblies are necessary for this test
RBF, TBF, SCAN, MVP, MSTE, Monitor
MSTE
CPU
Reset
ROM
RAM
Super
I/O
VRAM
AGDC
IDE
CTL
COM1
Foot S/W
Parallel Port
COM2
MOD
Quasi--RS232C
19200 BPS
RTC
From NOVA_BUS
SYSTEM DRAM TEST
ILLUSTRATION 2--18