GE MEDICAL SYSTEMS
LOGIQ
a
200 PROPRIETARY MANUAL
2138854
DIAGNOSTICS
2–13
REV 2
2–3–2 System Diagnosis (Continued)
TABLE 2–2
DIAGNOSIS TEST DETAIL
RESULT CODE
Meaning
Passed
Checked
Skipped
The DRAM test, at booting, is skipped.
No PIOP
ECG Module is absent.
211
HV test is failed.
311
Front/End path test is failed.
312
Dynamic range test is failed.
321
B Mode interpolation test is failed.
322
B Mode coordination test is failed.
323
B Mode ALM test is failed.
324
B Mode XY memory is failed.
330
The first Cine memory test is failed or CINE option is not installed
331
The Second Cine memory test is failed.
332
The Third Cine memory test is failed.
336
The First Cine memory or the Second Cine memory test is failed.
337
The Second Cine memory or the Third Cine memory test is failed.
341
M Mode Communication test is failed.
342
M Mode XY memory test is failed.
343
M Mode interpolation test is failed.
344
M Mode path test is failed.
411
Battery test is failed.
412
SCSI test is failed.
413
SRAM test is failed.
414
Real Time Clock test is failed.
415
Counter is failed.
416
DRAM test is failed.
417
RS232C test is failed.
418
Application ROM checksum is failed.
419
VRAM test is failed.
Содержание LOGIQ 200
Страница 4: ......
Страница 8: ......
Страница 10: ...05 23 00 MAC Page 2 of 2 ...
Страница 28: ...05 23 00 MAC Page 2 of 2 ...
Страница 87: ...LOGIQ α200 ...
Страница 88: ...LOGIQ 200 PRO ...
Страница 144: ......
Страница 168: ...GE MEDICAL SYSTEMS LOGIQ 200 PRO Series PROPRIETARY MANUAL 2242594 DIAGNOSTICS 2 31 B MODE CINE TEST ILLUSTRATION 2 15 ...
Страница 190: ......
Страница 196: ......