Countermeasure
Content
SensCode
SensKey
Restart the ImageQuant LAS 4000 and Image-
Quant LAS 4000 Control Software
Lifting table error during
startup diagnosis
B001
3
Filter changer error during
startup diagnosis
B002
Focus error during startup
diagnosis
B004
Iris error during startup diag-
nosis
B008
Epi light source error during
startup diagnosis
B010
UV transluminator error dur-
ing startup diagnosis
B020
White light table error during
startup diagnosis
B040
A combination of errors B004
to B020 above are generated
during startup diagnosis
B0
**
except described
above
Lifting table error
B101
Filter changer error
B102
Focus error
B104
Iris error
B108
Transmitted light source error
B110
UV transilluminator error
B120
White light table error
B140
Errors B104 to B120 above
are generated in combina-
tion.
B1
**
except described
above
Restart the ImageQuant LAS 4000 and Image-
Quant LAS 4000 Control Software
Communication error be-
tween camera head and Im-
ageQuant LAS 4000
A000
4
Communication retry error
between camera head and
ImageQuant LAS 4000
A100
A200
70
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
6 Troubleshooting
6.5 Message “ImageQuant LAS 4000 hardware error occurred"