3 Installation
This chapter provides information regarding installation of the ImageQuant LAS 4000.
3.1 Site requirements
The following table gives a summary of power supply and environmental requirements. For detailed
information, refer to Section 7.1 Main specifications, on page 79.
Requirement
Parameter
13 A, 100 - 120 V or 10 A, 200 - 240 V
Power supply
50 to 60 Hz
Line frequency
15°C to 28°C, fluctuation < 10°C/h
Ambient temperature
(Storage: -25°C to 70°C)
Stable laboratory bench, load capacity > 100kg/m
2
Placement
At least 20 cm free on all sides
30% to 70%, non-condensing
Humidity
(Storage: 5% to 100%, non-condensing)
3.2 Transport
The ImageQuant LAS 4000 weighs approximately 57 kg and requires at least two people to lift and
move it.
Before moving the instrument:
1 Turn off the instrument.
2 Disconnect the power cord.
3 Remove the lens, transmitted light source and any loose parts from inside of the IDX. Refer to the
ImageQuant LAS 4000 User Manual.
3.3 Unpacking
When unpacking:
• Check the equipment for any apparent damage before installing.
26
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
3 Installation