Environmental conditions
Operating temperature/humidity conditions
(with temperature fluctuation below 10ºC
per hour or lower)
:15°C to 28°C
Temperature
(no dew condensation)
:30 % to 70 % RH
Humidity
When the above conditions cannot be satisfied, modify the facilities accordingly.
Transportation/storage conditions
: -25°C to 70°C
Temperature
: 5 % to 100 % RH (no dew condensation)
Humidity
Installation location conditions
1 Do not install the equipment in an area where the temperature varies widely.
2 Do not install the equipment near a source of heat.
3 Do not install the equipment in an area where it may get wet or flooded.
4 Do not install the equipment in an area where it may be exposed to corrosive gas.
5 Do not install the equipment in a dusty area.
6 Do not install the equipment in an area constantly or excessively exposed to os-
cillations or impacts.
7 Do not install the equipment in an area exposed to direct sunlight.
: Indoors
Operation site
: 2000 m or lower
Maximum operating altitude
: Transient overvoltage
Overvoltage category category II
: Pollution Degree 2
Rated pollution applied
USB 2.0
Do not connect the ImageQuant LAS 4000 USB connector to a computer not certified
with UL60950-1 (UL listed) and IEC60950-1.
Analysis unit interface
Noise : 70 dB (A) or lower
Others
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
83
7 Reference information
7.1 Main specifications