Countermeasure
Meaning
Error message
Place the NP tray at tray posi-
tion 3. Use the LAS High Sens.
lens
The NP tray can be used
only with the LAS High
Sens. lens and at tray posi-
tion 3
NP Tray can only be used in tray po-
sition 3, with the FUJINON VRF43LMD
lens and the chemiluminescence
method.
NP tray is not in the proper position.
Insert the tray and ensure it is
pushed fully in
The tray is not detected
Please set the tray so that contact is
made with the sensor at the rear of
the cabinet.
Start button - during exposure
Countermeasure
Meaning
Error message
Change the settings appropri-
ately, or perform an exposure
in Manual mode
The automatic exposure
function is unable to calcu-
late a correct exposure
time
Cannot continue with the automatic
exposure function. The sample may
be out of the photometric range, or
a setting may need to be adjusted.
Reduce the total exposure time
so that the UV transilluminator
is lit for less than 10 minutes
The UV transilluminator
was turned off
Exposure has been stopped.
The UV illuminator was turned off
because the 10-minute limit was
reached.
The image may be incorrectly
exposed. Perform the exposure
again.
The door was opened dur-
ing exposure
Attention! The door was opened
during exposure.
Start button - after exposure
Countermeasure
Meaning
Error message
Press OK to discard the
recorded image(s), or Cancel to
go back and save the images
No images were saved
No image file was saved. Do you
want to continue?
Choose an application to
launch in Launch Application
in Preference in the Edit menu
Failed to launch the exter-
nal application. The desig-
nated file cannot be found.
The save and launch function failed.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
75
6 Troubleshooting
6.6 Error messages and countermeasures