Countermeasure
Meaning
Error message
Change the method and/or the
tray position in the
Method/Tray Position dialog
The default settings have
been changed
Components in ImageQuant LAS
4000 have changed.
Please change the exposure method.
Please change settings in
Method/Tray Position window.
Start button - starting exposure
Countermeasure
Meaning
Error message
Enter an appropriate time value
The time value is inappro-
priate
The interval time must be greater
that 10 seconds when using the In-
crement/Repetition mode.
The maximum exposure time when
using the EtBr method is 10 minutes.
The maximum exposure time when
using the UV light is 10 minutes.
The maximum interval time is 2
hours.
The maximum exposure time is 30
hours.
The maximum interval time when
using the EtBr method is 60 seconds.
The maximum interval time when
using the UV light is 60 seconds.
Input a valid filename in the
Program Settings dialog
No file name is given in
Program Settings
Please input the file name using the
program settings window.
Choose a suitable exposure
time
The exposure time is not
appropriate
Please input a valid exposure time.
Change the sensitivity setting,
or expose an image in Manual
mode
In the Digitize Method, the
Auto function cannot be
used when the sensitivity
is High, Super or Ultra
Cannot start the exposure.
Automatic exposure is only available
when using standard sensivity in the
digitization method.
Automatic exposure cannot be used
at the sensitivity or resolution.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
73
6 Troubleshooting
6.6 Error messages and countermeasures