Countermeasure
Meaning
Error message
Perform a flat frame calibration
The CCD has not been cali-
brated
Cannot start flat frame calibration
because the CCD calibration for High
Resolution mode does not exist.
Please run CCD calibration first.
Option menu - CCD Calibration
Countermeasure
Meaning
Error message
Perform the CCD calibration
again
The door was opened dur-
ing exposure, so the image
may not have been ex-
posed correctly
The door was opened during expo-
sure. This image may be inaccurate.
Option menu - CCD Calibration Pro
Countermeasure
Meaning
Error message
Use the CCD Calibration option
The FlatFrame Calibration
Pro function cannot be
used in chemilumescent
mode
Cannot use the flat frame calibration
for the following settings: Light none,
Filter No1, FUJINON VRF43LMD lens
Change the settings appropri-
ately, or perform an exposure
in Manual mode
The automatic exposure
function is unable to calcu-
late a correct exposure
time
Cannot continue with the automatic
exposure function. The sample may
be out of the photometric range, or
a setting may need to be adjusted.
Help menu
Countermeasure
Meaning
Error message
Ensure that the documentation
and PDF reader software is in-
stalled
The PDF documents can-
not be opened
Failed to open the User Manual.
Please note that a PDF reader (e.g.
Adobe Reader) is needed.
Failed to open End-User Licence
Agreement. Please note that a PDF
reader (e.g. Adobe Reader) is needed.
Failed to open the Getting Started
guide. Please note that a PDF reader
(e.g. Adobe reader) is needed.
Getting Started with ImageQuant LAS 4000 28-9607-59 AB
77
6 Troubleshooting
6.6 Error messages and countermeasures