Version No.: 1
SYSTEM COMPAX 40E
Job Card IST 001
1 of
4
Purpose:
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
Date: February 10, 1997
Time:
Personnel:
1
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
83
SECTION 1
SPARES REQUIRED
MX100 Tube
SECTION 2
EQUIPMENT REQUIRED
Protective pad, metal rod.
SECTION 3
SAFETY PRECAUTIONS
None
SECTION 4
PRELIMINARY STEPS
None
SECTION 5
INSTALLATION PROCEDURES
The following procedure describes the steps for replacing an MX100 Tube.
NEVER REMOVE THE ROD DURING X
−
RAY TUBE (OR COLLIMATOR)
REPLACEMENT UNTIL STEP 8. OF THIS PROCEDURE
Follow these instructions carefully and read the procedure through before
starting work.