System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
29
AEC for Table Bucky Technique
AEC for Wall Bucky Technique
AEC for Tomo Technique
Step
Description
Done
Supporting Documents
Í
Í
From previous page
ÍÍ
ÍÍ
See SM MPH 50/65/80, X
−
ray unit, Dir.
2137354
−
100, Chapter 1, Installation, Job
Card RG005.
2 People
6 Hours
AEC CALIBRATION
FILM PROCESSING
Processor ready and operational
Film/Screen characteristics defined
Film Image Quality available
ÍÍ
ÍÍ
NOTE: Film processor must be
fully operational at this stage to
proceed with AEC calibration.
Image quality tools (sensitometer,
etc.) must also be available.
CALIBRATION
A36
A37
Continues on next page
SID SCALES
POSITIONING
XT Suspension
RS 85 Tubestand
SG60/SG100
XT:
Refer to SM 2118733
−
100 Dir
46
−
019803, section 8.6
or Dir 2118910
−
100, section 12.6
RS
−
85:
Refer to Interface SM
2165897
−
100 (Chapter 1, section 4 and 5)
SG60:
Refer to SM 2150293
−
100
SG100:
Refer to SM 2150297
−
100
Chapter 1, section 9
A38
1 Person
1 Hour