Version No.: 1
SYSTEM COMPAX 40E
Job Card IST 001
1 of
4
Purpose:
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
Date: February 10, 1997
Time:
Personnel:
1
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
75
SECTION 1
SPARES REQUIRED
SECTION 2
EQUIPMENT REQUIRED
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PRELIMINARY STEPS
SECTION 5
INSTALLATION PROCEDURES
Room light, X
−
Ray light and Wall Bucky are generally powered with:
. 220 VAC in EUROPE
. 110 VAC in USA
Refer to this Jobcard section 5.2 to select the proper setting.
For DOOR OPEN SWITCH and EXPOSURE ON SIGNAL LAMP connection, refer to MPH A6 A4
Wall Bucky Board schematic, Illustration 7 of this Job Card.