Version No.: 2
SYSTEM COMPAX 40E
Job Card IST 001
1 of
4
Purpose:
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
Date: August 7, 1997
Time:
Personnel:
1
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
63
SECTION 1
SPARES REQUIRED
SECTION 2
EQUIPMENT REQUIRED
SECTION 3
SAFETY PRECAUTIONS
SECTION 4
PRELIMINARY STEPS
WITCH THE POWER OFF ON XT SUSPENSION BEFORE CHANGING TAP
SELECTION ON XT CONTROL BOARD.
SECTION 5
INSTALLATION PROCEDURES
5.1
Configuration During HLA
For XT Suspension installation in Europe the Collimator lamp voltage is configured by connecting
an electrical jumper to J10 (24V
AC
) for the Philips lamp or J11 (26V
AC)
for the GE type lamp, when
a 230VAC input line is on XT transformer (refer to System Inspection Report and Illustration 1).
Note:
An XT Suspension integrated in the US is furnished with jumper connected to J9.