Job Card IST 001
2 of
4
X−RAY TUBE & COLLIMATOR
INSTALLATION ON XT SUSPENSION
System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
84
1. Place a protective pad on the foot end of the Table top.
2. Insert the metal rod (refer to Illustration 10) onto the Column to immobilize the Tube Arm
vertically.
3. Immobilize the RS 85 Column carriage so it cannot move longitudinally.
4. Rotate the Tube Assy 180
°
to ”Tube Output Up” to remove the Collimator.
5. Remove the Collimator and set it aside on the pad. Cable disconnect should not be
necessary.
6. Remove MX100 Tube and replace it with the new tube.
7. Re
−
install the Collimator onto the X
−
ray Tube assy.
8. Remove the rod from the RS 85 Column to free the Tube Arm.
ILLUSTRATION
10
METAL ROD ON VERTICAL CARRIAGE
METAL ROD
WARNING
Fig.1