Section 2.0 - Regulatory Clearances
Section 2.0
Regulatory Clearances
2.1
Regulatory Clearances
MINIMUM CLEARANCES UNDER U.S. FEDERAL REGULATIONS AND NATIONAL
STANDARDS: 29 CFR 1910 (OSHA), NFPA 70E (STANDARD FOR ELECTRICAL
SAFETY IN THE WORKPLACE), AND NFPA 101 (LIFE SAFETY CODE):
is a map of clearance requirements for U.S. regulatory compliance. See clearance tables
on the following pages for detailed dimensional clearances. Please note all systems installed in the
United States must comply with all Federal and local regulations. For installations outside the
United States, country-specific or other local regulatory clearance requirements must be met.
Figure 4-1 Regulatory Clearance Requirements for BrightSpeed System Configurations
Note:
See
914 to 1219mm
(36 to 48 in.)
914 to 1219mm
(36 to 48 in.)
Table
Super
Structure
Footprint
Gantry Super
Structure
Footprint
Table
Base
Cradle
&
Extender
Foot
Switch
Asm
3212mm (10ft. 6in.)
1861mm (6ft. 1in. )
36
"
28
“
36
"
36"
2057 mm
( 6ft. 9in. )
Regulatory Compliance
Workspace Clearance
Perpendicular to Equipment
Regulatory Compliance 711 mm
28 inch Egress Clearance
Reference Regulatory Clearance Tables
on following pages
for minimum and variable distance definitions
and additional metric conversions.
28 “
A
C B
A) Minimum Room
(Less than 711 mm (28 in. ))
B) Minimum Room
(Less than 914 mm (36 in. ))
C) Recommended Room
(914 mm (36 in.) or Greater)
NGPDU
914 to 1219mm
(36 to 48 in.)
Depth
762mm(30in.)
Width
1981mm(78in.)
Height
914 mm
(36 in.)
711 mm
(28 in.)
914 mm
(36 in.)
914 mm
(36 in.)
914 mm
(36 in.)
711 mm
(28 in.)
610 mm
(24 in.)