Chapter 4 - Room Planning
Page 67
nning
•
Cradle Movement limitation can be set at any position.
•
Table Height Limitation can not be set (no change for footprint at Gantry Front side).
•
Scannable range is depends on the Gantry Rear space (distance to the wall), but need to
consider the Service Clearance and country's local regulation for Gantry Rear space.
4.2
Instruction of using Short Footprint function
NOTICE
Cradle limitation must comply with country or local regulatory clearance requirements.
Cradle movement limitation set for short footprint must be approved by customer during pre-
installation.
1.) Refer to
, use floor template with ruler to prearrange the layout and calculate the
cradle scannable limitation (A).
2.) Make GE siting print to meet regulatory and service clearance requirements.
3.) Record the distance from cradle limitation to wall (X) and cradle scannable limitation (A) for
installation.
A
C
D
B
F
711
mm
(28
in.)
E
914 mm
(36
in.)
914
mm
(36 in.)
643
mm
(25in.)
3212
mm
(10ft. 6in.)
Cradle
&
Extender
X
A=643
mm
+B-X