Page 86
Section 8.0 - Structural Requirements
8.5.3
Equipment Location
To minimize the interference, the CT equipment should be placed on a solid floor, located as far as
possible from the following vibration sources:
8.6
Walls: Scan Window
The recommended patient viewing window dimensions are 1219 mm wide by 1067 mm high (48 in.
x 42 in. ). The location of the window is dependent on the position of Operator Workspace position.
Consult
of this chapter and a
qualified radiological health physicist
for radiation
protection requirements of glass (lead content and thickness).
Note:
The operator at the Operator Workspace must be able to view the patient during a scan.
8.7
Ceiling Requirement
The weight of "Boom in a room LCD monitor" assy is 50 kg., and ceiling shall have enough strength
against the specification.
•
Parking lots
•
Roadways
•
Subways
•
Trains
•
Hallways
•
Elevators
•
Heliports
•
Hospital power plants containing pumps, motors, air handling
equipment and air conditioning units