Page 152
Section 6.0 - UPS Interconnect
Section 6.0
UPS Interconnect
Figure 9-6 Typical UPS Interconnect
NOTICE
UPS Kit B7999ZA REQUIRES installation of one of the A1 Panels listed below
.
Table 9-16 BrightSpeed Series Partial UPS Back-up Options
Note:
Conduit is required between:
•
A1 and UPS
•
UPS and PDU
•
PDU and A1
37’ (45’)
GE-supplied
Input cable - 7’ (15')
A1
SEO
UPS
3 phase
power source
Output cable - 7’ (15’)
35’ max.
Legend
Cable lengths are: Usable (Actual)
37’ (45’)
PDU
Customer-supplied
PDU Type &
Model #
Max. Mom.
kVA Rating
Recommended Main Disconnect (A1) Cat #
Optional
Partial UPS Kit
Cat #
America
EMEA
Asia
NGPDU-3
2326492-3
90kVA
E4502AB (90A) or
E4502AC (110A)
(incl. Auto Restart &
Integrated UPS
Control)
E46001AC
(incl. Auto Restart &
Integrated UPS
Control)
E4502AC (110A)
(incl. Auto Restart &
Integrated UPS
Control)
B7999ZA
alt. E4502KY
(includes 5169128
9155-10GE model
10KVA, 2ph UPS &
hardware kit)
REQUIRES one of
the A1 Panels shown
at left