Section 1.0 - Alternate Cover Removal Options
1.3
Rear Cover Removal
Rather than storing the rear cover straight back from the gantry, the cover can be moved and
stored on the right or left side or angled if there is space available while still maintaining service
access to the gantry. Additionally, the cover can be moved to the side of the table or out of the
scan room to a temporary storage
location.
For rooms with a surface floor duct (without ramps) behind the gantry, the rear cover cannot be
moved to the side of the gantry. Due to the weight of the gantry cover, lifting it over a surface floor
duct without ramps is prohibited.
1.4
Condition References
There are three possible minimum service space requirements based on the construction of the
wall directly adjacent to the side of the gantry. The following three conditions determine the
minimum space requirement that would apply to the room based on the special conditions of the
wall:
•
Condition 1
If the side of the system being serviced is directly facing an ungrounded surface or wall without
live voltage panels and without surface mounted ducts or conduits the minimum space
requirement is 914 mm (36 in).
•
Condition 2
If side of the system being serviced is directly facing a grounded surface or wall the minimum
space requirement 1067 mm (42 in).
•
Condition 3
If side of the system being serviced is directly facing a surface or wall with live voltage panels,
surface mounted ducts, or conduits the minimum space requirement is 1219 mm (48 in).