4 Application wizards
4.5 Assay wizards
64
Biacore T100 Software Handbook BR-1006-48 Edition AE
Step 5. Control samples
Specify the details of control samples and the frequency with which they should
be run. You may choose not to run control samples at all: however, including
control samples is generally recommended as an aid in assessing the
performance of the assay.
If you check only
Run control samples
, the controls will be run once only at the
start of the assay. If you check
Repeat control samples every…
, the controls will
be run after the first calibration curve and then at the specified interval for
sample cycles (not counting repeated calibration curves), and once at the end
of the assay. Use the
Cycle Run List
option in the
System Preparations
step to
verify where control samples will be run during the assay.
Control samples are specified in terms of sample ID and expected
concentration. The expected concentrations should lie within the range covered
by the calibration curve.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......