Biacore T100 Software Handbook BR-1006-48 Edition AE
45
Application wizards 4
Step 2. Injection parameters
Enter the name of the ligand to be tested and the contact time and flow rate.
Recommended settings are a contact time of 120 seconds at 5 or 10 µl/min: you
may need to use a longer contact time if preconcentration of ligand on the
sensor surface proves to be slow.
The surface is washed with a “regeneration” injection at the end of each cycle to
remove any ligand that might remain on the surface. The recommended
solution for this procedure is 50 mM NaOH.
Step 3. System preparations
See Section 4.2.5. Run immobilization pH scouting at the same temperature as
you intend to run the immobilization (electrostatic preconcentration is however
usually fairly insensitive to temperature).
Step 4. Rack positions
See Section 4.2.6. Immobilization pH scouting requires one position for ligand
solution at each pH tested and one for the surface wash solution. Accept or
change the rack positions for the various solutions required (see Section 4.2.6).
Step 5. Prepare run protocol
Edit the
Prepare Run Protocol
text if desired (Section 4.2.7). This text will be
displayed at the start of a run when the wizard template is used. Save the wizard
template if required and start the run.
pH scouting results
When the wizard run is completed, the results are opened automatically in the
Evaluation Software, with an overlay plot of the sensorgrams adjusted to the
start of the sample injection. The same overlay plot is created when a saved run
from
Immobilization pH Scouting
is opened in the Evaluation Software.
Note:
Opening a saved run in the Control Software does not generate the
overlay plot automatically.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......