Biacore T100 Software Handbook BR-1006-48 Edition AE
219
Method examples and recommendations Appendix B
B.3
GST kinetics
The recommended method for using a GST-tagged ligand is set up with
Dual
detection in
General Settings
. Detection settings
Multi
and
Single
are not
recommended. The method involves 5 conditioning cycles with injection of
recombinant GST (provided in the GST Capture Kit from GE Healthcare) over both
flow cells followed by regeneration:
The sample analysis cycle uses two capture injections, one to capture
recombinant GST on the reference surface only (flow path setting
First
) and the
second to capture GST-tagged ligand on the active surface only (flow path
setting
Second
). A stabilization time of 180 seconds is included after capturing
the ligand on the active surface, to allow for dissociation of any weakly bound
components. The sample is then injected over both surfaces. This approach
provides a reference surface that mimics the active surface with respect to
occupancy of the anti-GST capturing sites.
Содержание Biacore T100
Страница 1: ...GE Healthcare Biacore T100 Software Handbook ...
Страница 2: ......
Страница 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Страница 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 24: ...2 Control Software general features 2 4 File storage 24 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 76: ...4 Application wizards 4 6 Control experiment wizards 76 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Страница 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 138: ...7 Data presentation tools 7 4 Report point table 138 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Страница 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 214: ...Appendix A Data import and export A 2 Importing data 214 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Страница 237: ......