26
•
Check whether the values in the fields
#1
and
#2
correlate with the values on the back of the calibration model.
Adopt the values of the calibration model.
•
Click on
Start
.
•
After the start, you can no longer interrupt the process. The scanner is calibrated. A message will inform you
about this process:
After successful completion of calibration, the message
3D calibration has been executed successfully
will be
displayed:
•
Click
OK
.
The dialog closes. You can now start scanning.