TR-CL180 Lens Controller - User Manual
Use the
p
and
q
buttons to set the
Error Detect flag:
F E
is enabled,
F-E
is
disabled. Press the
SEL
button.
Switched output is set up.
11.0.4 Setting up analog control
Press and hold
SEL
for 1 second and follow the keystrokes below:
CH1
is displayed. Press the
SEL
button.
Use the
p
and
q
buttons to select
SAn
(Other options displayed are
SAc
and
SCo
). Press the
SEL
button.
Use the
p
and
q
buttons to set the
Error Detect flag:
F E
is enabled,
F-E
is
disabled. Press the
SEL
button
Analog control is set up.
11.0.5 Setting numerical values
When the TR-CL180 requires you to enter a numeric value, the right hand
digit flashes to indicate you can use the
p
and
q
keys to change the
value.
The optical power can be set from -99 to +99.9, but depending on the
value, the resolution changes as follows:
Optical power value
Number format
Step value
-99 to -10
-99
1
-9.9 to -0.1
-9.9
0.1
0 to 9.99
9.9
0.01
10 to 99.9
99.9
0.1
—
20
—