Cell Connections
40
The cell cable connections used in
Save Both Half Cells
mode are shown in Table 5-2. The orange Counter
Sense lead
must
be connected.
Table 5-2
Cell Cable Connections for Three-Electrode Battery Testing
Color
Type
Name
Normal Connection
Blue
Banana Plug
Working Sense
Connect to positive battery terminal
Green
Banana Plug
Working Electrode
Connect to positive battery terminal
White
Pin Jack
Reference
Connect to the reference electrode
Red
Banana Plug
Counter Electrode
Connect to negative battery terminal
Orange
Banana Plug
Counter Sense
Connect to negative battery terminal
Black
Pin Jack
Floating Ground
Leave open or connect to a Faraday shield
The voltage of the full cell is measured between the counter sense and the working sense leads.
The white pin jack on the cell cable is normally connected to a reference electrode. The potential between this
lead and the Working Sense lead is reported as the positive half-cell potential.
ZRA Mode Cell Connections
The Interface 5000 can function as a precision Zero Resistance Ammeter (ZRA). In this mode it maintains two
metal samples at the same potential and measures the current flow between the samples. It can also measure
the potential of the samples versus a reference electrode. The cell cable connections for ZRA mode are shown
in Table 5-3. The connections are very similar to those for the potentiostat and galvanostat modes.
In ZRA mode a second working electrode is substituted for the counter electrode, and the Orange Counter
Sense lead
must
be connected.
Table 5-3
Cell Cable Connections for ZRA Mode
Color
Type
Name
Normal Connection
Blue
Banana Plug
Working Sense
Connect to metal sample #1
Green
Banana Plug
Working Electrode
Connect to metal sample #1
White
Pin Jack
Reference
Connect to a reference electrode
Red
Banana Plug
Counter Electrode
Connect to metal sample #2
Orange
Banana Plug
Counter Sense
Connect to metal sample #2
Black
Pin Jack
Floating Ground
Leave open or connect to a Faraday shield
The counter sense and the working sense lead are each connected to different metal samples. In ZRA mode,
the Interface 5000 is normally programmed to maintain zero volts between these leads. It therefore maintains
the two metal samples at the same voltage.
The white pin jack on the cell cable is normally connected to a reference electrode. The potential between this
lead and the working sense lead is reported as the cell potential.
Содержание Interface 5000
Страница 2: ...2 ...
Страница 18: ...Introduction 18 ...
Страница 30: ...Installation 30 ...
Страница 36: ...Calibration 36 ...
Страница 64: ...Appendix A Interface 5000 Specifications 64 ...
Страница 72: ...Appendix D Power LED Blink Codes and Error Messages 72 ...
Страница 74: ...Appendix E CE Certificate Declaration of Conformity 74 Low Voltage Certificate of Conformance ...
Страница 75: ...Appendix E CE Certificate Declaration of Conformity 75 RFI Certificate of Conformance ...
Страница 76: ...Appendix E CE Certificate Declaration of Conformity 76 ...
Страница 80: ...Appendix F Heat in Interface 5000 Multichannel Systems Interface Power Hub Systems 80 ...
Страница 85: ......
Страница 86: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com ...