Panel Indicators and Connectors
48
Cell On LED
The
Cell On
LED is to the left of the Front Panel Cell connector.
The
Cell On
LED glows yellow whenever the Interface 5000 is actively applying voltage or current to the
electrochemical cell attached to the Cell Cable. Avoid touching the cell cable leads whenever the
Cell On
LED
is lit, because the quality of the data being collected in your experiment may be compromised.
If you need to make changes to your cell leads, do so between experiments, when the
Cell On
LED is off and
the potentiostat is inactive.
In a typical experimental sequence, the
Cell On
LED is off between experiments and during any open-circuit
potential measurements. It glows yellow whenever the cell is polarized.
Overload LED
The
Overload
LED is just below the
Cell On
LED.
The
Overload
LED is normally unlit. When it glows red, some circuit in the Interface 5000 has exceeded its
normal operating limit. Conditions that generate Overloads include:
•
The absolute value of the differential electrometer output voltage (the difference in voltage between
the Working and Reference leads) exceeds
±
6.125 V. This condition is known as an
E Overload
.
The
Cell On
LED does not indicate a dangerous condition when it is lighted. The voltages
generated by the Interface 5000 are generally considered safe. Still, avoid touching the cell leads when the
cell is on.
Содержание Interface 5000
Страница 2: ...2 ...
Страница 18: ...Introduction 18 ...
Страница 30: ...Installation 30 ...
Страница 36: ...Calibration 36 ...
Страница 64: ...Appendix A Interface 5000 Specifications 64 ...
Страница 72: ...Appendix D Power LED Blink Codes and Error Messages 72 ...
Страница 74: ...Appendix E CE Certificate Declaration of Conformity 74 Low Voltage Certificate of Conformance ...
Страница 75: ...Appendix E CE Certificate Declaration of Conformity 75 RFI Certificate of Conformance ...
Страница 76: ...Appendix E CE Certificate Declaration of Conformity 76 ...
Страница 80: ...Appendix F Heat in Interface 5000 Multichannel Systems Interface Power Hub Systems 80 ...
Страница 85: ......
Страница 86: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com ...