Appendix B: Interface 5000 Cell Connectors
66
Table B-2
Sense Connector
Pin(s)
Signal Name
Use
1
CBL_ID2
One of four cable ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
2
CBL_ID1
One of four cable ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
3
Ground
The potentiostat’s floating ground. Can be used to shield the cell if very low
currents need to be measured. Also used as a shield for the counter electrode
cable.
4
Counter-
sense Shield
The shield for the counter-sense electrode input. Driven to the same
potential as Pin 11. Left open at the cell end of the cell cable.
5,6,12,14
Reference
Shield
The shield for the reference electrode input. Driven to the same potential as
Pin 13. Left open at the cell end of the cell cable.
7,15
Work Sense
Shield
The shield for the work sense electrode input. Driven to the same potential
as Pin 8. Left open at the end of the cell cable.
8
Work Sense
Connected to the working electrode in most cases (see Chapter 5). This lead
has a 261 Ω resistor in the cell end of th
e cable. Custom cell cables are likely
to require a similar resistor.
9
CBL_ID3
One of four cable ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
10
CBL_ID0
One of four cable ID bits. Used to identify the type of cell cable attached to
the unit. Pull to a logic High through a resistor. Ground to set the bit low.
11
Counter
Sense
Connected to the counter electrode in ZRA-mode and stack-mode
experiments (see Chapter
5). This lead has a 261 Ω resistor in the cell end of
the cable. Custom cell cables are likely to require a similar resistor.
13
Reference
Electrode
Connected to the reference electrode in most cases (see Chapter 4
Error!
Reference source not found.
). This lead has a 261 Ω resistor in the cell end
of the cable. Custom cell cables are likely to require a similar resistor.
Содержание Interface 5000
Страница 2: ...2 ...
Страница 18: ...Introduction 18 ...
Страница 30: ...Installation 30 ...
Страница 36: ...Calibration 36 ...
Страница 64: ...Appendix A Interface 5000 Specifications 64 ...
Страница 72: ...Appendix D Power LED Blink Codes and Error Messages 72 ...
Страница 74: ...Appendix E CE Certificate Declaration of Conformity 74 Low Voltage Certificate of Conformance ...
Страница 75: ...Appendix E CE Certificate Declaration of Conformity 75 RFI Certificate of Conformance ...
Страница 76: ...Appendix E CE Certificate Declaration of Conformity 76 ...
Страница 80: ...Appendix F Heat in Interface 5000 Multichannel Systems Interface Power Hub Systems 80 ...
Страница 85: ......
Страница 86: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com ...