Appendix C: I/O Connectors
69
Table C-2
Monitor Connector
Pin
Name
Use
1
Signal Ground
A high-quality floating ground. Signals are relative to this ground
2
Signal Ground
Another high-quality floating ground. Signals are relative to this
ground
3
RTD Positive
The Positive RTD Input. 1 mA flows between this pin and Pin 4.
4
RTD Negative
The Negative RTD input. Con
nected to the Interface 5000’s floating
ground.
5
E_sig
E Monitor Signal,
±
6.5536 V full scale
6
EXT_SIG
External Signal In,
±
6.5536 V full scale
7
I_Sig
I_Monitor Signal,
±
3.2678 V full scale
8
Aux ADC In
–
The minus input of the differential Aux ADC Input
9
Aux ADC In +
The positive input of the differential Aux ADC Input
I Monitor Signal
The I Monitor signal represents the output of the Interface 5000 current measurement circuit. It can be treated
as a two-wire differential signal with the negative side connected to the Interface 5000 Signal Ground. With the
exception of the filtering described below, it is the raw signal with no offset or gain applied. IE Stability
capacitors slow the response.
Scaling on this signal is
3 V for
the nominal full scale current on the selected current range. Cathodic currents
cause a positive output voltage. If the software is auto-ranging the current-range selection, this signal is
discontinuous at each range change.
The I Monitor signal is filtered using an RLC circuit. It has a bandwidth of approximately 350 kHz when
connected to a high-impedance input.
E Monitor Signal
The E Monitor signal is derived from the output of the Interface 5
000’s differential electrometer circuit.
It can
be treated as a two-wire differential signal with the negative side connected to the Interface 5000 Signal
Ground. With the exception of the filtering described below, it is the raw voltage signal, with no offset or gain
applied. With the exception of the filtering described below, it is buffered representation of the voltage
difference between the white and blue cell cable leads.
Caution:
The negative side of the E Monitor signal is connected to the Interface 5000
Floating Ground. Connection of signal to an earth-ground-referenced apparatus can compromise the
Interface 5
000’s ability to float
, and invalidate data collected on earth-grounded cells.
Caution:
The ground side of the I Monitor signal is connected to the Interface 5000
Floating Ground. Connection signal to an earth-ground-referenced apparatus can compromise the Interface
5
000’s ability to float
, and invalidate data collected on earth-grounded cells.
Содержание Interface 5000
Страница 2: ...2 ...
Страница 18: ...Introduction 18 ...
Страница 30: ...Installation 30 ...
Страница 36: ...Calibration 36 ...
Страница 64: ...Appendix A Interface 5000 Specifications 64 ...
Страница 72: ...Appendix D Power LED Blink Codes and Error Messages 72 ...
Страница 74: ...Appendix E CE Certificate Declaration of Conformity 74 Low Voltage Certificate of Conformance ...
Страница 75: ...Appendix E CE Certificate Declaration of Conformity 75 RFI Certificate of Conformance ...
Страница 76: ...Appendix E CE Certificate Declaration of Conformity 76 ...
Страница 80: ...Appendix F Heat in Interface 5000 Multichannel Systems Interface Power Hub Systems 80 ...
Страница 85: ......
Страница 86: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com ...