Safety Considerations
12
Be careful when operating the Interface 5000 in an enclosed space (such as an enclosed relay rack or NEMA
enclosure). The temperature within the enclosure must not exceed 45
C. You may need to provide ventilation
holes or even forced air-cooling for the enclosed space if excessive temperature rise occurs.
Defects and Abnormal Stresses
Treat your Interface 5000 as potentially hazardous if any of the following is true of the unit:
•
It shows visible damage,
•
It does not operate properly,
•
It has been stored for a long period of time under unfavorable conditions,
•
It has been dropped or subjected to severe transport stress,
•
It has been subjected to environmental stress (corrosive atmosphere, fire, etc.).
Do not use your Interface 5000 or any other apparatus if you think it could be hazardous. Have it checked by
qualified service personnel.
Environmental Limits
Note that there are environmental limit conditions on the storage, shipping and operation of this equipment.
The Interface 5000 has
not
been designed for outdoor use.
Storage
Ambient temperature
–
40
C to 75
C
Relative humidity
Maximum 90% non-condensing
Shipping
Same as storage plus
Acceleration
Maximum 30 G
Operation
Ambient temperature
0
C to 45
C
Relative humidity
Maximum 90% non-condensing
Cleaning
Disconnect the Interface 5000 from all power sources prior to cleaning.
Use a cloth
lightly
dampened with either clean water or water containing a mild detergent, to clean the outside
of the Interface 5000 enclosure. Alternatively, you may use isopropyl alcohol. Do not use a wet rag or allow
Warning:
The Interface 5000 is not designed for operation in conditions where liquid water
may enter the chassis, or water vapor may condense within the chassis. Operation of an Interface 5000 that
has water within the chassis can create a safety hazard, which could result in personal injury or death.
Caution:
Do not block the airflow into or out of the Interface 5000 chassis. The circuitry is
thermally protected so over-temperature will cause the instrument to enter a shut-down state. If shut-down
occurs during data-acquisition, experimental data could be lost. Running the Interface 5000 without
adequate cooling could also shorten the time to failure of some of the circuitry.
Содержание Interface 5000
Страница 2: ...2 ...
Страница 18: ...Introduction 18 ...
Страница 30: ...Installation 30 ...
Страница 36: ...Calibration 36 ...
Страница 64: ...Appendix A Interface 5000 Specifications 64 ...
Страница 72: ...Appendix D Power LED Blink Codes and Error Messages 72 ...
Страница 74: ...Appendix E CE Certificate Declaration of Conformity 74 Low Voltage Certificate of Conformance ...
Страница 75: ...Appendix E CE Certificate Declaration of Conformity 75 RFI Certificate of Conformance ...
Страница 76: ...Appendix E CE Certificate Declaration of Conformity 76 ...
Страница 80: ...Appendix F Heat in Interface 5000 Multichannel Systems Interface Power Hub Systems 80 ...
Страница 85: ......
Страница 86: ...734 Louis Drive Warminster PA 18974 Tel 215 682 9330 Fax 215 682 9331 www gamry com info gamry com ...