Error

WHAt doES It MEAN

WHAt CAN I do?

 

 

 

Under Pressure the Control Unit 
can’t reach its target maximum 
compression. this often indicates 
that there is a leak in the pneumatic 
compression circuit, either in the 
connector hose or the Wrap itself. 
this is a Soft Error. there is no 
danger to the horse being treated 
or the Control Unit, and therapy will 
continue.

once treatment has finished, try the same 
pressure with a different Wrap, if one is 
available. When you have determined if the 
problem is isolated to one Wrap or not, call 
Game ready Equine customer service at  
888.GAMErEAdY (426-3732) You can 
continue to use your system without harming 
the Control Unit. You can even continue using 
the affected Wrap, but this error will persist. 

 
 

 

low Flow Warning the Control Unit 
has detected a temporary interruption 
of fluid flow, but it was able to 
correct the situation by reducing the 
compression setting. this is a Soft 
Error. there is no danger to the horse 
being treated or the Control Unit, and 
therapy will continue. 

this error can often be corrected by 
reapplying the Wrap so that it is snug against 
the body part being treated with no major 
folds or creases. Please consult your Wrap 
use guide for more detailed instruction. If the 
problem persists, call customer service 
at 888.GAMErEAdY (426-3732) 

 

 

 

over Pressure Alarm the Control 
Unit has detected excessive pressure 
in the air chamber. In order to prevent 
any possible harm to the horse being 
treated, the unit will stop therapy.  
this is a Hard Error. the unit needs 
to be turned off and restarted. 

• Turn the Control Unit off and back on. 
• Apply the same Wrap more tightly. 
• Apply a different Wrap. 

If the problem persists, customer service at 
888.GAMErEAdY (426-3732) 

 
 

 

Pressure too High Alarm  
the Control Unit has detected that  
the Wrap has not properly deflated.  
In order to prevent any possible harm 
to the horse being treated, the unit  
will stop therapy. this is a Hard Error. 
the unit needs to be turned off and 
restarted. 

this error often comes about when the horse 
being treated moves around during treatment 
and puts pressure on the air bladder. If this is 
not the case try the following: 

• Turn the Control Unit off and back on. 
• Apply the same Wrap more tightly. 
• Apply a different Wrap. 

If the problem persists, call customer service 
at 888.GAMErEAdY (426-3732)  

 
 

low Flow Alarm the Control Unit  
has detected an interruption of fluid 
flow that it could not correct. In order 
to prevent possible damage to the  
fluid pump, the unit will stop therapy. 
this is a hard Error. the unit needs 
to be turned off and restarted. 

• Check the hose and Wrap for kinks that   

 

   might cut off fluid flow. 

• Make sure there is water in the ice box. 
• Make sure that the ice box filter is clean. 
• Reapply the Wrap so that it is snug against 

the horse. 
If the problem persists, call customer service 
at 888.GAMErEAdY (426-3732) 

dry Pump Alarm the Control Unit has 
detected a dry pump. this error is 
displayed after one attempt to correct 
the problem has failed. In order to 
prevent possible damage to the fluid 
pump, the unit will stop therapy. 
this is a Hard Error. the unit needs to 
be turned off and restarted.

• Make sure there is water in the ice box. 
• Verify that the ice box filter is not clogged. 
• Turn the Control Unit off and on again. 
• Select “No Pressure” and run the system

 

   for 1 minute using the largest Wrap   
   available, lying it on the floor or table next 
   to the Control Unit (not on the  horse).  
If the problem persists, call customer s

ervice 

at 888.GAMErEAdY (426-3732)

t r o U B l E S H o o t I N G

Содержание Equine 550250

Страница 1: ...EQU INE Contr ol Uni t User s Manual Model 550250...

Страница 2: ......

Страница 3: ...in order to receive warranty service Who We Are CoolSystems was founded by a pioneer in NASA spacesuit technology who began working with athletes athletic trainers and doctors to expand the technolog...

Страница 4: ......

Страница 5: ...__________________ 12 Operating The System In The Bag _____________________________ 14 The AC Adapter How To Use The AC Adapter ____________________________________ 14 Care And Storage _______________...

Страница 6: ...the System and disconnect the power line cord from its electrical outlet before performing any maintenance procedures Caution To avoid the risk of electrical shock do not remove any panels from the Co...

Страница 7: ...cted to the Ault Model MW128 power supply and Class II ungrounded when connected to the GlobTek Model GTM21097 5012 power supply Degree of safety in the presence of flammable anesthetics or oxygen 5 5...

Страница 8: ...optional Connector Hose Heat Exchanger C O N TROL U N I T OR I E N T A T I O N S Figure A Upright Figure B Side The Control Unit can sit upright Fig A or on one side Fig B to fit on a shelf Caution Do...

Страница 9: ...n priming the Full Leg Heat Exchanger the Control Unit alarm will sound after 45 seconds the low flow icon will flash and the Control Unit will stop This is because there isn t enough water returning...

Страница 10: ...on its side place it back in its upright position before opening the ice box lid by pulling up on the latch Press the power button to turn off the unit Unplug unit from the AC adapter and empty excess...

Страница 11: ...st open wounds sores rashes infections or stitches If skin has been breached first apply a sterile dressing over the wound then apply the Game Ready Equine Wrap After treatment remove the Wrap from th...

Страница 12: ...o another part of the horse be sure that the air bladder of the Wrap is completely deflated before starting another treatment on a new area of the horse To deflate the Wrap simply unplug the Wrap from...

Страница 13: ...panel to adjust the level of compression A corresponding icon will appear on the display panel as shown on page 7 Note The pressure level cannot be adjusted while the system is in use To adjust the p...

Страница 14: ...sure To adjust the pressure setting during a treatment you must first press the pause button see page 13 for details General Pressure Guidelines Acute Trauma No Pressure Bowed Tendons Low Pressure Sus...

Страница 15: ...Press the C F button to choose your desired mode With a full ice box typically the Control Unit will provide cold therapy between 35 F and 50 F Setting Treatment Time Each time the power button is us...

Страница 16: ...e the Control Unit from the bag to empty the ice box O P E R A T I N G T H E S Y S T E M I N T H E B A G How to Use the AC Adapter The Game Ready Equine System operates using external power through th...

Страница 17: ...Equine Control Unit is equipped with a special filter designed to keep dirt and debris out of the fluid system To assure consistent flow of water through the system the filter should be checked and c...

Страница 18: ...f the Wrap If the ice box is sealed shut hydrogen peroxide can form a gas that could build excess pressure inside the unit and potentially damage the system Wrap Storage Do not store Wraps in extreme...

Страница 19: ...adhesive Handle Grip Screw Plug Logo Medallion Mains power quality should be that of a typical commercial or hospital environment Surge protection is recommended to protect Control Unit If the user o...

Страница 20: ...he problem persists customer service at 888 GAMEREADY 426 3732 Pressure Too High Alarm The Control Unit has detected that the Wrap has not properly deflated In order to prevent any possible harm to th...

Страница 21: ...problem at startup This is a Hard Error The unit needs to be turned off and restarted Disconnect the Wrap from the Control Unit Turn the Control Unit off and on again without a Wrap connected If the p...

Страница 22: ...way and doing so will cause water to leak through its air vent While a minimal amount of water leakage should not damage the system immediately place the unit in its proper position upright or on its...

Страница 23: ...s property upon receipt The replacement Game Ready Control Unit is warranted under this written warranty and is subject to the same limitations and exclusions for the remainder of the original warrant...

Страница 24: ...ine Customer Service Department 1201 Marina Village Pkwy Suite 200 Alameda CA 94501 510 559 3940 Toll free phone number 1 888 GAMEREADY 1 888 426 3732 www gamereadyequine com Game Ready Equine Control...

Отзывы: