t r o U B l E S H o o t I N G

Error

WHAt doES It MEAN

WHAt CAN I do?

 

 

 

Pressure Sensor Alarm  
the Control Unit has detected a 
problem calibrating the pneumatic 
compression circuit on startup.  
this is a Hard Error. the Unit needs  
to be turned off and restarted. 

• Disconnect the Wrap from the Control Unit. 
• Turn the Control Unit off and on again 

    without a Wrap connected. 
this should clear the error. If the  
problem persists, call customer service 
at 888.GAMErEAdY (426-3732) 
 

 
 

 

Systems Error the Control Unit has 
determined that the fluid pump may 
be working too hard. this could be 
caused by ice or debris in fluid circuit. 
In order to prevent possible damage 
to the fluid pump, the unit will stop 
therapy. this is a Hard Error. the unit 
needs to be turned off and restarted.

• Turn the Control Unit off and back on again. 
• Check the filter. 
• If that does not solve the problem turn the 

 

   Control Unit off for 20 minutes before 
   turning it on again to resume use. 
If the problem persists, call customer service 
at 888.GAMErEAdY (426-3732) 

 

 

 

Self-test Error the Control Unit has 
detected an electronic problem at 
startup. this is a Hard Error. the unit 
needs to be turned off and restarted. 

• Disconnect the Wrap from the Control Unit. 
• Turn the Control Unit off and on again 

 

    without a Wrap connected. 
If the problem persists, call customer service 
at 888.GAMErEAdY (426-3732)

 
 

 

low Battery Warning the Control 
Unit has detected a voltage below 
10V. the unit will continue to run, but 
performance may drop. this is a Soft 
Error. there is no danger to the horse 
being treated or the Control Unit, and 
therapy will continue. 

• If you are using a rechargeable battery 

 

    pack, you should recharge the battery  
    very soon. 

• If you are using an alkaline battery pack, 

 

    you should change the batteries soon. 

• If you are not using a battery pack, you may 

 

    be connected to a poor source of power or  
    you may be using an incorrect AC Adapter. 

 
 

low Battery Alarm the Control Unit 
has detected a voltage below 9.5V. 
the unit will not be able to run 
below this point. this is a Hard Error. 
the unit needs to be turned off and 
restarted. 

• If you are using a rechargeable battery 

 

    pack, you need to recharge the battery. 

• If you are using an alkaline battery pack, 

 

    you need to change the batteries. 

• If you are not using a battery pack, you may 

 

    be connected to a poor source of power or  
    you may be using an incorrect AC Adapter. 

Содержание Equine 550250

Страница 1: ...EQU INE Contr ol Uni t User s Manual Model 550250...

Страница 2: ......

Страница 3: ...in order to receive warranty service Who We Are CoolSystems was founded by a pioneer in NASA spacesuit technology who began working with athletes athletic trainers and doctors to expand the technolog...

Страница 4: ......

Страница 5: ...__________________ 12 Operating The System In The Bag _____________________________ 14 The AC Adapter How To Use The AC Adapter ____________________________________ 14 Care And Storage _______________...

Страница 6: ...the System and disconnect the power line cord from its electrical outlet before performing any maintenance procedures Caution To avoid the risk of electrical shock do not remove any panels from the Co...

Страница 7: ...cted to the Ault Model MW128 power supply and Class II ungrounded when connected to the GlobTek Model GTM21097 5012 power supply Degree of safety in the presence of flammable anesthetics or oxygen 5 5...

Страница 8: ...optional Connector Hose Heat Exchanger C O N TROL U N I T OR I E N T A T I O N S Figure A Upright Figure B Side The Control Unit can sit upright Fig A or on one side Fig B to fit on a shelf Caution Do...

Страница 9: ...n priming the Full Leg Heat Exchanger the Control Unit alarm will sound after 45 seconds the low flow icon will flash and the Control Unit will stop This is because there isn t enough water returning...

Страница 10: ...on its side place it back in its upright position before opening the ice box lid by pulling up on the latch Press the power button to turn off the unit Unplug unit from the AC adapter and empty excess...

Страница 11: ...st open wounds sores rashes infections or stitches If skin has been breached first apply a sterile dressing over the wound then apply the Game Ready Equine Wrap After treatment remove the Wrap from th...

Страница 12: ...o another part of the horse be sure that the air bladder of the Wrap is completely deflated before starting another treatment on a new area of the horse To deflate the Wrap simply unplug the Wrap from...

Страница 13: ...panel to adjust the level of compression A corresponding icon will appear on the display panel as shown on page 7 Note The pressure level cannot be adjusted while the system is in use To adjust the p...

Страница 14: ...sure To adjust the pressure setting during a treatment you must first press the pause button see page 13 for details General Pressure Guidelines Acute Trauma No Pressure Bowed Tendons Low Pressure Sus...

Страница 15: ...Press the C F button to choose your desired mode With a full ice box typically the Control Unit will provide cold therapy between 35 F and 50 F Setting Treatment Time Each time the power button is us...

Страница 16: ...e the Control Unit from the bag to empty the ice box O P E R A T I N G T H E S Y S T E M I N T H E B A G How to Use the AC Adapter The Game Ready Equine System operates using external power through th...

Страница 17: ...Equine Control Unit is equipped with a special filter designed to keep dirt and debris out of the fluid system To assure consistent flow of water through the system the filter should be checked and c...

Страница 18: ...f the Wrap If the ice box is sealed shut hydrogen peroxide can form a gas that could build excess pressure inside the unit and potentially damage the system Wrap Storage Do not store Wraps in extreme...

Страница 19: ...adhesive Handle Grip Screw Plug Logo Medallion Mains power quality should be that of a typical commercial or hospital environment Surge protection is recommended to protect Control Unit If the user o...

Страница 20: ...he problem persists customer service at 888 GAMEREADY 426 3732 Pressure Too High Alarm The Control Unit has detected that the Wrap has not properly deflated In order to prevent any possible harm to th...

Страница 21: ...problem at startup This is a Hard Error The unit needs to be turned off and restarted Disconnect the Wrap from the Control Unit Turn the Control Unit off and on again without a Wrap connected If the p...

Страница 22: ...way and doing so will cause water to leak through its air vent While a minimal amount of water leakage should not damage the system immediately place the unit in its proper position upright or on its...

Страница 23: ...s property upon receipt The replacement Game Ready Control Unit is warranted under this written warranty and is subject to the same limitations and exclusions for the remainder of the original warrant...

Страница 24: ...ine Customer Service Department 1201 Marina Village Pkwy Suite 200 Alameda CA 94501 510 559 3940 Toll free phone number 1 888 GAMEREADY 1 888 426 3732 www gamereadyequine com Game Ready Equine Control...

Отзывы: