19
t r o U B l E S H o o t I N G
Error
WHAt doES It MEAN
WHAt CAN I do?
Pressure Sensor Alarm
the Control Unit has detected a
problem calibrating the pneumatic
compression circuit on startup.
this is a Hard Error. the Unit needs
to be turned off and restarted.
• Disconnect the Wrap from the Control Unit.
• Turn the Control Unit off and on again
without a Wrap connected.
this should clear the error. If the
problem persists, call customer service
at 888.GAMErEAdY (426-3732)
Systems Error the Control Unit has
determined that the fluid pump may
be working too hard. this could be
caused by ice or debris in fluid circuit.
In order to prevent possible damage
to the fluid pump, the unit will stop
therapy. this is a Hard Error. the unit
needs to be turned off and restarted.
• Turn the Control Unit off and back on again.
• Check the filter.
• If that does not solve the problem turn the
Control Unit off for 20 minutes before
turning it on again to resume use.
If the problem persists, call customer service
at 888.GAMErEAdY (426-3732)
Self-test Error the Control Unit has
detected an electronic problem at
startup. this is a Hard Error. the unit
needs to be turned off and restarted.
• Disconnect the Wrap from the Control Unit.
• Turn the Control Unit off and on again
without a Wrap connected.
If the problem persists, call customer service
at 888.GAMErEAdY (426-3732)
low Battery Warning the Control
Unit has detected a voltage below
10V. the unit will continue to run, but
performance may drop. this is a Soft
Error. there is no danger to the horse
being treated or the Control Unit, and
therapy will continue.
• If you are using a rechargeable battery
pack, you should recharge the battery
very soon.
• If you are using an alkaline battery pack,
you should change the batteries soon.
• If you are not using a battery pack, you may
be connected to a poor source of power or
you may be using an incorrect AC Adapter.
low Battery Alarm the Control Unit
has detected a voltage below 9.5V.
the unit will not be able to run
below this point. this is a Hard Error.
the unit needs to be turned off and
restarted.
• If you are using a rechargeable battery
pack, you need to recharge the battery.
• If you are using an alkaline battery pack,
you need to change the batteries.
• If you are not using a battery pack, you may
be connected to a poor source of power or
you may be using an incorrect AC Adapter.