System will not turn on: 
1.  Check that the AC adapter is securely plugged into a working electrical outlet. 
2.  If using the Battery Pack, check that batteries are not dead. 

Control Unit turns off during use: 
If the system current draw rises above 1.99 amps during operation, the Control Unit will shut off to prevent 
damage to the pumps. Usually this is an indicator of a blockage in the fluid or air flow. Check the Wrap and 
the connector hose to be sure there are no kinks, creases or folds. Unplug the Control Unit to reset the system, 
and try the Wrap again. If the Control Unit shuts off again, there is either a problem with the Wrap or the 
Control Unit. try a different Wrap to determine if it is the Control Unit or the Wrap and call Game ready 
customer service at 1-888-GAMErEAdY for further assistance. 

Control Unit leaks water when placed on its side: 
1. If the Control Unit is on its side, check to be sure that the control panel is not upside down. the unit is  
  not designed to rest this way, and doing so will cause water to leak through its air vent. While a minimal 
  amount of water leakage should not damage the system, immediately place the unit in its proper position: 
  upright or on its side with the Control Unit right side up. 
2.  If there is excess water on the rubber gasket around the mouth of the ice box, the lid will not seal properly, 
  and some of the water is likely to drip out when placed on its side. Simply wipe off excess water from the 
  gasket and close the lid securely. 
3  Check the ice box gasket to be sure it’s seated properly around the lip of the ice box. If it is not flat in the 
  groove, push it down so that it is flush with the lip of the ice box. If the gasket is damaged, call Game 
  ready customer service at 1-888-GAMErEAdY for further assistance. 

leak at the Wrap – hose connection point: 
If the connector leaks water where the hose plugs into the Wrap it is probably because the Wrap was plugged 
in at a severe enough angle to cause the connector to not seal properly. to fix the problem, simply unplug the 
Wrap and plug it back in being careful to plug it in straight. If the problem persists call Customer Service at 
1-888-GAMErEAdY. 

 

t r o U B l E S H o o t I N G

Содержание Equine 550250

Страница 1: ...EQU INE Contr ol Uni t User s Manual Model 550250...

Страница 2: ......

Страница 3: ...in order to receive warranty service Who We Are CoolSystems was founded by a pioneer in NASA spacesuit technology who began working with athletes athletic trainers and doctors to expand the technolog...

Страница 4: ......

Страница 5: ...__________________ 12 Operating The System In The Bag _____________________________ 14 The AC Adapter How To Use The AC Adapter ____________________________________ 14 Care And Storage _______________...

Страница 6: ...the System and disconnect the power line cord from its electrical outlet before performing any maintenance procedures Caution To avoid the risk of electrical shock do not remove any panels from the Co...

Страница 7: ...cted to the Ault Model MW128 power supply and Class II ungrounded when connected to the GlobTek Model GTM21097 5012 power supply Degree of safety in the presence of flammable anesthetics or oxygen 5 5...

Страница 8: ...optional Connector Hose Heat Exchanger C O N TROL U N I T OR I E N T A T I O N S Figure A Upright Figure B Side The Control Unit can sit upright Fig A or on one side Fig B to fit on a shelf Caution Do...

Страница 9: ...n priming the Full Leg Heat Exchanger the Control Unit alarm will sound after 45 seconds the low flow icon will flash and the Control Unit will stop This is because there isn t enough water returning...

Страница 10: ...on its side place it back in its upright position before opening the ice box lid by pulling up on the latch Press the power button to turn off the unit Unplug unit from the AC adapter and empty excess...

Страница 11: ...st open wounds sores rashes infections or stitches If skin has been breached first apply a sterile dressing over the wound then apply the Game Ready Equine Wrap After treatment remove the Wrap from th...

Страница 12: ...o another part of the horse be sure that the air bladder of the Wrap is completely deflated before starting another treatment on a new area of the horse To deflate the Wrap simply unplug the Wrap from...

Страница 13: ...panel to adjust the level of compression A corresponding icon will appear on the display panel as shown on page 7 Note The pressure level cannot be adjusted while the system is in use To adjust the p...

Страница 14: ...sure To adjust the pressure setting during a treatment you must first press the pause button see page 13 for details General Pressure Guidelines Acute Trauma No Pressure Bowed Tendons Low Pressure Sus...

Страница 15: ...Press the C F button to choose your desired mode With a full ice box typically the Control Unit will provide cold therapy between 35 F and 50 F Setting Treatment Time Each time the power button is us...

Страница 16: ...e the Control Unit from the bag to empty the ice box O P E R A T I N G T H E S Y S T E M I N T H E B A G How to Use the AC Adapter The Game Ready Equine System operates using external power through th...

Страница 17: ...Equine Control Unit is equipped with a special filter designed to keep dirt and debris out of the fluid system To assure consistent flow of water through the system the filter should be checked and c...

Страница 18: ...f the Wrap If the ice box is sealed shut hydrogen peroxide can form a gas that could build excess pressure inside the unit and potentially damage the system Wrap Storage Do not store Wraps in extreme...

Страница 19: ...adhesive Handle Grip Screw Plug Logo Medallion Mains power quality should be that of a typical commercial or hospital environment Surge protection is recommended to protect Control Unit If the user o...

Страница 20: ...he problem persists customer service at 888 GAMEREADY 426 3732 Pressure Too High Alarm The Control Unit has detected that the Wrap has not properly deflated In order to prevent any possible harm to th...

Страница 21: ...problem at startup This is a Hard Error The unit needs to be turned off and restarted Disconnect the Wrap from the Control Unit Turn the Control Unit off and on again without a Wrap connected If the p...

Страница 22: ...way and doing so will cause water to leak through its air vent While a minimal amount of water leakage should not damage the system immediately place the unit in its proper position upright or on its...

Страница 23: ...s property upon receipt The replacement Game Ready Control Unit is warranted under this written warranty and is subject to the same limitations and exclusions for the remainder of the original warrant...

Страница 24: ...ine Customer Service Department 1201 Marina Village Pkwy Suite 200 Alameda CA 94501 510 559 3940 Toll free phone number 1 888 GAMEREADY 1 888 426 3732 www gamereadyequine com Game Ready Equine Control...

Отзывы: