background image

15

 

C A r E   A N d   S t o r A G E 

Cleaning 
the Game ready Equine System is a valuable part of your equipment. Please handle the Control Unit as 
you would a laptop computer. When you are finished using the system for the day, unplug the Control Unit 
from the electrical outlet. disconnect both the Wrap and AC adapter from the Control Unit. Empty the 
water and ice from the ice box. the interior of the ice box should be cleaned once every three months. 
(See Page 16) to clean the exterior of the Control Unit, simply wipe with a soft cloth rinsed in warm water 
and a mild disinfectant
containing any of the following agents:  
A. Mild detergent  
B. rubbing alcohol  
to clean the internal fluid circuit of the Control Unit fill the ice box 3-4 inches deep with hydrogen  
peroxide, attach any Wrap and run the unit for 2 minutes with no pressure. 
Caution: leave lid open when using hydrogen peroxide. 

Storage and transportation Environmental Conditions:
temperature: 1° - 50° C (33° - 122° F)
relative Humidity: 10% - 95% non-condensing
Store the Game ready Equine System in the carrying bag or on a shelf, preferably with the lid open  
to allow the interior to dry.
Caution: do not keep in extreme hot or cold temperatures (above 120°F or below 33°F).  
Avoid leaving in the trunk of a hot or freezing car. 

Cleaning and replacing the Filter 
Your Game ready Equine Control Unit is equipped with a special filter designed to keep dirt and debris out 
of the fluid system. to assure consistent flow of water through the system, the filter should be checked and 
cleaned every three months of regular use. If the filter becomes clogged, the Control Unit alarm will beep, 
the  
”low flow“ icon will flash, and the unit will go into pause mode. 
Caution: do not run the Control Unit without the filter. doing so may cause damage to the unit  
and will void your warranty. 
In addition to cleaning the filter every three months, it should be replaced after one year of regular usage. 
Call us toll-free at 888-GAMErEAdY for a free replacement filter. 

to clean or replace the filter, follow these steps: 
1. turn off power and unplug the system from its electrical outlet. 
2.  With the Control Unit in the upright position, open the ice box lid. 
3.  Empty any water and ice in the ice box. 
4.  locate the filter inside the ice box with your hand,  
  as shown in Fig. A. 
5. Press the silver metal button on top of the filter, as shown  
  in Fig. B. You should hear a click as the filter is released. 
6. Pull the filter straight out from its housing 
  (Note: the release button is part of the housing and 
  therefore remains attached to the inside of the unit). 
7. Clean the filter by rinsing it thoroughly under 
  warm running water. Shake it a few times to clear 
  out any remaining debris. 
8. re-insert the filter in the unit by reaching in and snapping  
  it back into the housing. You should hear it click into place. 

Figure A

Figure B

Содержание Equine 550250

Страница 1: ...EQU INE Contr ol Uni t User s Manual Model 550250...

Страница 2: ......

Страница 3: ...in order to receive warranty service Who We Are CoolSystems was founded by a pioneer in NASA spacesuit technology who began working with athletes athletic trainers and doctors to expand the technolog...

Страница 4: ......

Страница 5: ...__________________ 12 Operating The System In The Bag _____________________________ 14 The AC Adapter How To Use The AC Adapter ____________________________________ 14 Care And Storage _______________...

Страница 6: ...the System and disconnect the power line cord from its electrical outlet before performing any maintenance procedures Caution To avoid the risk of electrical shock do not remove any panels from the Co...

Страница 7: ...cted to the Ault Model MW128 power supply and Class II ungrounded when connected to the GlobTek Model GTM21097 5012 power supply Degree of safety in the presence of flammable anesthetics or oxygen 5 5...

Страница 8: ...optional Connector Hose Heat Exchanger C O N TROL U N I T OR I E N T A T I O N S Figure A Upright Figure B Side The Control Unit can sit upright Fig A or on one side Fig B to fit on a shelf Caution Do...

Страница 9: ...n priming the Full Leg Heat Exchanger the Control Unit alarm will sound after 45 seconds the low flow icon will flash and the Control Unit will stop This is because there isn t enough water returning...

Страница 10: ...on its side place it back in its upright position before opening the ice box lid by pulling up on the latch Press the power button to turn off the unit Unplug unit from the AC adapter and empty excess...

Страница 11: ...st open wounds sores rashes infections or stitches If skin has been breached first apply a sterile dressing over the wound then apply the Game Ready Equine Wrap After treatment remove the Wrap from th...

Страница 12: ...o another part of the horse be sure that the air bladder of the Wrap is completely deflated before starting another treatment on a new area of the horse To deflate the Wrap simply unplug the Wrap from...

Страница 13: ...panel to adjust the level of compression A corresponding icon will appear on the display panel as shown on page 7 Note The pressure level cannot be adjusted while the system is in use To adjust the p...

Страница 14: ...sure To adjust the pressure setting during a treatment you must first press the pause button see page 13 for details General Pressure Guidelines Acute Trauma No Pressure Bowed Tendons Low Pressure Sus...

Страница 15: ...Press the C F button to choose your desired mode With a full ice box typically the Control Unit will provide cold therapy between 35 F and 50 F Setting Treatment Time Each time the power button is us...

Страница 16: ...e the Control Unit from the bag to empty the ice box O P E R A T I N G T H E S Y S T E M I N T H E B A G How to Use the AC Adapter The Game Ready Equine System operates using external power through th...

Страница 17: ...Equine Control Unit is equipped with a special filter designed to keep dirt and debris out of the fluid system To assure consistent flow of water through the system the filter should be checked and c...

Страница 18: ...f the Wrap If the ice box is sealed shut hydrogen peroxide can form a gas that could build excess pressure inside the unit and potentially damage the system Wrap Storage Do not store Wraps in extreme...

Страница 19: ...adhesive Handle Grip Screw Plug Logo Medallion Mains power quality should be that of a typical commercial or hospital environment Surge protection is recommended to protect Control Unit If the user o...

Страница 20: ...he problem persists customer service at 888 GAMEREADY 426 3732 Pressure Too High Alarm The Control Unit has detected that the Wrap has not properly deflated In order to prevent any possible harm to th...

Страница 21: ...problem at startup This is a Hard Error The unit needs to be turned off and restarted Disconnect the Wrap from the Control Unit Turn the Control Unit off and on again without a Wrap connected If the p...

Страница 22: ...way and doing so will cause water to leak through its air vent While a minimal amount of water leakage should not damage the system immediately place the unit in its proper position upright or on its...

Страница 23: ...s property upon receipt The replacement Game Ready Control Unit is warranted under this written warranty and is subject to the same limitations and exclusions for the remainder of the original warrant...

Страница 24: ...ine Customer Service Department 1201 Marina Village Pkwy Suite 200 Alameda CA 94501 510 559 3940 Toll free phone number 1 888 GAMEREADY 1 888 426 3732 www gamereadyequine com Game Ready Equine Control...

Отзывы: