5
Protection against electric shock (5.1):
the Game ready power supply (Ault Model MW128) is considered to be Class I (grounded) equipment.
the Game ready power supply (Globtek Model GtM21097-5012) is considered to be Class II
(ungrounded) equipment.
the Game ready unit is considered to be Class II (ungrounded) equipment.
Protection against harmful ingress of water (5.3):
this product provides ordinary protection against ingress of water. the Game ready system is
considered to be Class I (ungrounded) when connected to the Ault Model MW128 power supply,
and Class II (ungrounded) when connected to the Globtek Model GtM21097-5012 power supply.
degree of safety in the presence of flammable anesthetics or oxygen (5.5):
Not suitable for use in an oxygen enriched environment or in the presence of flammable anesthetics.
Mode of operation (5.6):
the equipment operates under continuous operation and continuous operation with intermittent loading
if operating the system continuously without a full load of ice. Users should allow the system to rest for
at least 15 minutes between each 90-minute treatment.
Powered by direct Current
Attention: Consult accompanying documents before use.
U l C l A S S I F I C A t I o N
!
12V/ 1.2A