6:9
AK 95 / AK 95 S Service Manual
HCEN 9138 Rev. 10.2005
Calibrations
The calibration is a “two point calibration”. The
two points are 0.0 mS/cm (RO-water) and
14.0 mS/cm (with concentrate).
NOTE!
Make sure the +500ml/min restrictor is mounted
before calibration.
NOTE!
• Before conductivity calibration, Chemical
cleaning must be performed, one with Citric acid
and one with Hypochlorite.
1.
Enter the
Service
menu,
Calib, Transd
and select
Intern
.
Set the display at the conductivity trans-
ducer to be calibrated.
The possible cond. cell calibrations are:
• CONDUCTIVITY A TRANSDUCER
• CONDUCTIVITY B TRANSDUCER
• CONDUCTIVITY P TRANSDUCER
• CONDUCTIVITY ABP TRANSDUCERS
2.
Connect the Gambro reference instrument
to the dialyzer connectors.
Connect both the A- and B-pickup tubes to
the MM 95.
Set the temperature to 37,5°C (by pressing
TEMP key) and let the AK 95 S and the refer-
ence instrument stabilize for at least 30 min-
utes.
Make sure that the machine is not in bypass, i.e.
the fluid bypass button is not lit.
There must
be flow through the measuring cell.
3.
Press
to enter the calibration routine.
4.
Wait until the reading on the reference
instrument has stabilized at 0.0 mS/cm.
Set REF SET to 00.0 with Set knob.
Press “TEMP” and check that the tem-
perature is 37,5°C.
Press
to get back the conductivity
calibration display.
Press
to enter the value.
Conductivity Transducer calibration
5.
Place A-pickup tube in the concentrate
container. The AK 95 S will now regulate
conductivity to approx. 14 mS/cm. The
green fluid path must be lit.
6.
Let the conductivity stabilize for 20 min-
utes. (green fluid path)
Press “TEMP” to check that temperature
is 37,5°C.
Press
to get back the conductivity
calibration display.
7.
Set the “REF. SET” value on the AK 95 S
display to the same value shown on the
reference instrument.
8.
Press
to enter the value.
9.
The machine will now ask you to confirm
to enter sensitivity calibration.
Press
to confirm.
To exit: press
(will stop calibration).
10.
Press
START
UF
STOP
to load the calibration values
into the EEPROM.
All the transducers A, B and P are all calibrated
in this way.
Содержание AK 95
Страница 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Страница 5: ...This page is intentionally left blank ...
Страница 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Страница 7: ......
Страница 17: ...AK 95 AK 95 S Service Manual HCEN9133Rev 01 2004 1 10 Installation Guide This page is intentionally left blank ...
Страница 19: ...AK 95 AK 95 S Service Manual HCEN9133Rev 01 2004 1 12 Installation Guide This page is intentionally left blank ...
Страница 20: ......
Страница 64: ...Technical Data AK 95 HCEN9135 Rev 10 2005 3 1 Chapter 3 Technical data and specifications for AK 95 ...
Страница 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Страница 103: ......
Страница 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Страница 126: ......
Страница 151: ...AK 95 AK 95 S Service Manual HCEN9138Rev 10 2005 6 24 Calibrations This page is intentionally left blank ...
Страница 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Страница 153: ......
Страница 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Страница 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Страница 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Страница 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Страница 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Страница 173: ...HCEN9094Rev 10 2005 7 10 UF measuring unit 8 10 14 17 7 11 9 4 13 16 2 15 1 3 6 9 12 18 20 19 21 ...
Страница 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Страница 187: ...HCEN9094Rev 10 2005 7 16 Variable flow unit to Step driver board 2 2 4 5 3 1 8 7 4 11 6 9 10 13 14 15 12 ...
Страница 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Страница 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Страница 193: ...HCEN9094Rev 10 2005 7 19 Protective Covers 3 5 4 2 7 1 6 8 The figures do not correspond to their actual sizes ...
Страница 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Страница 203: ...HCEN9094Rev 10 2005 7 24 Front MM95 18 8 4 8 5 7 20 15 14 6 2 11 21 19 10 9 12 15 3 17 15 13 13 16 14 17 12 1 ...
Страница 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Страница 207: ...HCEN9094Rev 10 2005 7 26 Feeding pumps 15 18 14 16 10 7 10 19 2 6 3 4 17 5 7 12 20 13 9 11 14 18 8 ...
Страница 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Страница 217: ...HCEN9094Rev 10 2005 7 31 Remote panel Option 18 6 7 8 12 7 14 15 16 17 13 11 10 9 5 4 3 2 1 22 23 24 25 26 19 20 21 ...
Страница 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Страница 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Страница 224: ......
Страница 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Страница 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Страница 229: ...AK 95 AK95SHCEN12078Rev 10 2005 10 1 Replacements Replacements Replacement matrix 2 Changeactions 4 FM Flow Diagram 5 ...
Страница 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Страница 244: ...11 10 Electrical Safety Inspections for AK machines HCEN12190 Rev 05 2006 This page is intentionally left blank ...
Страница 247: ...AK 95 AK 95 S Service Manual ...