Disinfection and rinse
Chem disinfection
When Disinfection is selected by pressing the
CHEM DISINF
key, the information display
will show three alternatives of disinfection
solutions.
The disinfectant fluid will be taken up by the B-
tube and if the B tube is not taken out of the
machine and/or if the A-tube is taken out, there
will be an attention.
When the filling phase is finished the AK 95 S
will give an attention that requests you to place
the B-tube back in the monitor. When this is
done the disinfection solution is distributed in all
of the fluid path and the dwell time starts. As
soon as the dwell time is done, an automatic
rinse will start.
It is possible to end the Disinfection dwell
program before it is finished by pressing
RINSE/DRAIN
key.
It is not possible to go from chemical disinfection
mode to treatment mode without a rinse. In case
of a power failure the monitor will save recovery
data and continue the disinfection when the
power returns.
Heat disinfection
During the heat disinfection program the inlet
water is heated and rinsed through the fluid
unit. The machine is then drained.
It is also possible to run the heat disinfection in
combination with Citric Acid 20% or with a
CleanCart installed.
For heat disinfection a heat exchanger is a
requirement. For 230V version the exchanger
must have an efficiency of minimum 60%. 110V
and 100 V requires 76% efficiency.
To get a proper heat disinfection, the fluid must
exceed 80°C. The exchanger is not only necessar-
ily to achieve proper temperature but it is also
important for stabilization of the heater control
system.
Heat disinfection is selected with the
HEAT
DISINF
button. The remaining disinfection
time shows on the time display as soon the
disinfection starts. How many minutes a pro-
gram lasts is seen below. Total time for Disinfec-
tion Programs are estimated and may vary.
Disinfection
programs
Time
[min]
230 V 110 V 100 V
With
UFD
installed
Heat
35 39 49
Heat Citric acid 20% 52 57 67
Heat CleanCart 49 54 64
Without UFD
Heat
33 37 47
Heat Citric acid 20% 50 55 65
Heat CleanCart 46 51 61
Four minutes before completion of the heat-
disinfection, the program enters the drain phase
and the heat disinfection lamp is turned off. The
heat disinfection completes when the
RINSE/
DRAIN
key is flashing.
When pressing the flashing RINSE/DRAIN key
after a completed disinfection, the machine will
perform a functional check (Fch).
Rinse/Drain
Rinse mode will be started automatically after a
disinfection as described above or manually by
pressing the
RINSE/DRAIN
key. The display
shows the rinse time and allows you to change it
if you wish. The rinse program will run for the
set time. Both A- and B- pick-up tubes must be
placed in the monitor, otherwise there will be an
attention. During the final drain phase the air
intake valve,
AIVA
, is open. When rinse is
completed the RINSE/DRAIN
key will be flash-
ing.
The machine will perform a functional check
(Fch), when pressing the flashing RINSE/
DRAIN key after a completed disinfection.
There is a possibility to reset and preset the
chemical disinfection mode by enter the service
mode. For more information, see chapter 4,
Technicians guide.
Содержание AK 95
Страница 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Страница 5: ...This page is intentionally left blank ...
Страница 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Страница 7: ......
Страница 17: ...AK 95 AK 95 S Service Manual HCEN9133Rev 01 2004 1 10 Installation Guide This page is intentionally left blank ...
Страница 19: ...AK 95 AK 95 S Service Manual HCEN9133Rev 01 2004 1 12 Installation Guide This page is intentionally left blank ...
Страница 20: ......
Страница 64: ...Technical Data AK 95 HCEN9135 Rev 10 2005 3 1 Chapter 3 Technical data and specifications for AK 95 ...
Страница 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Страница 103: ......
Страница 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Страница 126: ......
Страница 151: ...AK 95 AK 95 S Service Manual HCEN9138Rev 10 2005 6 24 Calibrations This page is intentionally left blank ...
Страница 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Страница 153: ......
Страница 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Страница 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Страница 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Страница 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Страница 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Страница 173: ...HCEN9094Rev 10 2005 7 10 UF measuring unit 8 10 14 17 7 11 9 4 13 16 2 15 1 3 6 9 12 18 20 19 21 ...
Страница 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Страница 187: ...HCEN9094Rev 10 2005 7 16 Variable flow unit to Step driver board 2 2 4 5 3 1 8 7 4 11 6 9 10 13 14 15 12 ...
Страница 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Страница 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Страница 193: ...HCEN9094Rev 10 2005 7 19 Protective Covers 3 5 4 2 7 1 6 8 The figures do not correspond to their actual sizes ...
Страница 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Страница 203: ...HCEN9094Rev 10 2005 7 24 Front MM95 18 8 4 8 5 7 20 15 14 6 2 11 21 19 10 9 12 15 3 17 15 13 13 16 14 17 12 1 ...
Страница 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Страница 207: ...HCEN9094Rev 10 2005 7 26 Feeding pumps 15 18 14 16 10 7 10 19 2 6 3 4 17 5 7 12 20 13 9 11 14 18 8 ...
Страница 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Страница 217: ...HCEN9094Rev 10 2005 7 31 Remote panel Option 18 6 7 8 12 7 14 15 16 17 13 11 10 9 5 4 3 2 1 22 23 24 25 26 19 20 21 ...
Страница 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Страница 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Страница 224: ......
Страница 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Страница 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Страница 229: ...AK 95 AK95SHCEN12078Rev 10 2005 10 1 Replacements Replacements Replacement matrix 2 Changeactions 4 FM Flow Diagram 5 ...
Страница 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Страница 244: ...11 10 Electrical Safety Inspections for AK machines HCEN12190 Rev 05 2006 This page is intentionally left blank ...
Страница 247: ...AK 95 AK 95 S Service Manual ...