5:11
AK 95 / AK 95 S
HCEN12189 Rev.
10. 2006
Maintenance Manual
Base-kit
12.
Enter Service/Calibr/Transducers/
External/ Arterial pressure transducer.
12.1
Press "Select" to start calibration routine.
12.2
Make sure that nothing is connected to the
transducer (zero pressure)
12.3
Check zero level of the Arterial Pressure
Transducer.
Tolerance: ± 5 mmHg.
Calibrate if necessary (see chapter
calibration for details).
13.
Enter Service/Calibr/Transducers/
External/ Venous pressure transducer
CPU P.
13.1
Press "Select" to start calibration routine.
13.2
Make sure that nothing is connected to the
transducer (zero pressure).
13.3
Check zero level of the Venous Pressure
Transducer.
Tolerance: ± 5 mmHg.
Calibrate if necessary (see chapter
calibration for details).
13.4
Enter Service/Calibr/Transducers/
External/ Venous pressure transducer
CPU C.
13.5
Press "Select" to start calibration routine.
13.6
Make sure that nothing is connected to the
transducer (zero pressure).
13.7
Check zero level of the Venous Pressure
Transducer.
Tolerance: ± 5 mmHg.
Calibrate if necessary (see chapter
calibration for details).
14.
Enter Diagnose/BM/Pumps and check that
the blood pump stops when the cover is
opened.
15.
Start the blood pump and let it run at
lowest possible speed. Open the blood
pump cover and place a magnet (1) on the
sensor for the cover. Stop the blood pump
rotor by manually grabbing the handle (2).
Check that the power to the blood pump
motor is switched off within 2 seconds.
16.
Check the zero level of the three pressure
transducers:
- Degassing pressure transducer
- HPG pressure transducer
- Pressure Dialysis, PD
Note!
To reach atmospheric pressure in
the transducer housing, disconnect the
corresponding tube.
17.
If installed: Calibrate the pH-sensor with a
buffer solution according to the Calibration
manual.
18.
Remove the blood leak detector cover and
clean the inside of the blood leak housing.
Mount a blood leak calibration cover. Make
sure that the filter is in horizontal posi-
tion, start the AK 95 S.
Enter the Service menu, select calibration,
transducer, intern and blood leak detector.
Place A-pickup tube in concentrate contai-
ner.
Press "Select" to enter the blood leak
calibration routine.
Note! Let the machine stabilize in
conductivity and temperature (37.5 °C) a
few minutes.
Wait for green fluid path.
Check the blood leak detector limits 0 and
100 ±10.
Calibrate if necessary (see chapter
calibration for details).
19.
Drain the AK 95 S and replace the blood
leak calibration cover.
20.
Attach pressure reference instrument
before the heater. Use a silicon T.
Note!
Be careful when opening the fluid
tubes. Put a piece of paper around the tube
connector to avoid leakage inside the
monitor.
Содержание AK 95
Страница 1: ...AK 95 AK 95 S Service Manual A better way to better care ...
Страница 5: ...This page is intentionally left blank ...
Страница 6: ...Installation Guide AK 95 AK 95 S January 2004 ...
Страница 7: ......
Страница 17: ...AK 95 AK 95 S Service Manual HCEN9133Rev 01 2004 1 10 Installation Guide This page is intentionally left blank ...
Страница 19: ...AK 95 AK 95 S Service Manual HCEN9133Rev 01 2004 1 12 Installation Guide This page is intentionally left blank ...
Страница 20: ......
Страница 64: ...Technical Data AK 95 HCEN9135 Rev 10 2005 3 1 Chapter 3 Technical data and specifications for AK 95 ...
Страница 102: ...Maintenance Manual AK 95 AK 95 S October 2006 GAMBRO ...
Страница 103: ......
Страница 125: ...5 22 MaintenanceManual AK95 AK95SHCEN12189Rev 10 2006 This page is intentionally left blank ...
Страница 126: ......
Страница 151: ...AK 95 AK 95 S Service Manual HCEN9138Rev 10 2005 6 24 Calibrations This page is intentionally left blank ...
Страница 152: ...Spare Parts List AK 95 AK 95 S October 2005 ...
Страница 153: ......
Страница 159: ...HCEN9094Rev 10 2005 7 3 Operator s panel 9 1 3 4 5 6 5 7 5 2 8 6 6 8 ...
Страница 161: ...HCEN9094Rev 10 2005 7 4 Blood pump 10 2 8 3 5 9 4 7 1 6 11 12 ...
Страница 163: ...HCEN9094Rev 10 2005 7 5 Blood pump Motor unit 8 7 6 9 2 10 3 4 5 1 11 ...
Страница 165: ...HCEN9094Rev 10 2005 7 6 Heparin pump 1 2 3 4 5 6 8 7 ...
Страница 169: ...HCEN9094Rev 10 2005 7 8 Magnetic Clamp unit 1 2 3 7 8 12 16 4 5 6 9 10 11 17 18 14 13 19 15 ...
Страница 173: ...HCEN9094Rev 10 2005 7 10 UF measuring unit 8 10 14 17 7 11 9 4 13 16 2 15 1 3 6 9 12 18 20 19 21 ...
Страница 181: ...HCEN9094Rev 10 2005 7 13 Heater Flow guard 3 9 1 2 10 4 8 7 11 12 13 5 6 3 ...
Страница 187: ...HCEN9094Rev 10 2005 7 16 Variable flow unit to Step driver board 2 2 4 5 3 1 8 7 4 11 6 9 10 13 14 15 12 ...
Страница 189: ...HCEN9094Rev 10 2005 Protective Protective 7 17 Boards 7 1 2 9 6 5 4 3 8 ...
Страница 191: ...HCEN9094Rev 10 2005 1 7 18 Rear box 6 3 7 5 4 2 ...
Страница 193: ...HCEN9094Rev 10 2005 7 19 Protective Covers 3 5 4 2 7 1 6 8 The figures do not correspond to their actual sizes ...
Страница 195: ...HCEN9094Rev 10 2005 T12AH 250AC BACH BACO HALT 7 20 AC DC 2 4 5 1 6 7 8 9 3 ...
Страница 203: ...HCEN9094Rev 10 2005 7 24 Front MM95 18 8 4 8 5 7 20 15 14 6 2 11 21 19 10 9 12 15 3 17 15 13 13 16 14 17 12 1 ...
Страница 205: ...HCEN9094Rev 10 2005 7 25 MM I O board 3 2 9 4 6 7 8 5 1 10 ...
Страница 207: ...HCEN9094Rev 10 2005 7 26 Feeding pumps 15 18 14 16 10 7 10 19 2 6 3 4 17 5 7 12 20 13 9 11 14 18 8 ...
Страница 215: ...HCEN9094Rev 10 2005 7 30 UFD Option 4 5 7 17 12 9 6 18 1 6 11 16 8 10 6 13 14 3 2 6 19 15 ...
Страница 217: ...HCEN9094Rev 10 2005 7 31 Remote panel Option 18 6 7 8 12 7 14 15 16 17 13 11 10 9 5 4 3 2 1 22 23 24 25 26 19 20 21 ...
Страница 219: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Страница 221: ...HCEN9094Rev 10 2005 7 32 Boards and Cablings EN1107 10 Step driver board ...
Страница 224: ......
Страница 225: ...9 1 HCEN9160 Revision10 2005 Service Manual Symbols Chapter9 SymbolDescription ...
Страница 228: ...9 4 HCEN9160Revision10 2005 Service Manual Symbols Thispageisleftintentionallyblank ...
Страница 229: ...AK 95 AK95SHCEN12078Rev 10 2005 10 1 Replacements Replacements Replacement matrix 2 Changeactions 4 FM Flow Diagram 5 ...
Страница 234: ...Replacements 10 6 AK 95 AK95SHCEN12078Rev 10 2005 This page is intentionally left blank ...
Страница 244: ...11 10 Electrical Safety Inspections for AK machines HCEN12190 Rev 05 2006 This page is intentionally left blank ...
Страница 247: ...AK 95 AK 95 S Service Manual ...